User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx
39
7.7 Operation
The L-ORD consists of following components:
Spot size adjustment
The laser spot size can be adjusted from 50 to 500 µm with the
adjustment ring.
Locking screw
The L-ord is connected to the slit lamp via the pin of the tonometer
mounting base at the slit lamp. The locking screw is used for fixation.
Spot focus unit
Adjustment of the beam focus.
Horizontal alignment
The horizontal position of the beam can be adjusted via this
adjusting wheel.
Vertical Alignment
The vertical position of the beam can be adjusted via this adjusting
wheel.
Deflection mirror
The deflection mirror is used to bring the aiming beam as well as the
laser beam in coincide to the observation beam path of the slit lamp microscope.
Eye protection filter
The coated filter glasses protect the operator's eyes from the laser
beams. An additional eye protection filter for the slit lamp is not necessary.
Zoom optic
Spot size adjustment
50 µm bis 500 µm
Locking screw
Horizontal alignment
adjusting wheel
Spot focus unit
Deflection mirror
Vertical alignment
adjusting wheel
Eye safety filter