User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx
13
4.2 Eye safety
As a safety measure against direct or indirect laser radiation, it is necessary that
everyone in the room wear laser safety glasses. The treating doctor is protected from
radiation by the laser protection integrated (eye safety filter) into the slit lamp. Eye
protection appropriate to the patient must be guaranteed for the patient.
When using the laser, only those laser safety glasses that are designed for the
wavelength of 514 nm and on which both the CE mark and the appropriate protection
level are noted.
The following laser safety glasses from A.R.C. Laser GmbH are suitable for the
CLASSIC 514:
•
AS01003 for 514 nm: Protection level D LB6+ I LB 6 + RM LB8 / Optical density
OD8+
There are different types of laser safety glasses that also allow spectacle wearers to
wear a tight goggle on top of their own which are protective to all sides. In particular, the
scattered radiation that does not directly come from the front of the eye may present
risks resulting from internal reflections on the glasses. Therefore, we encourage you to
wear laser safety glasses which also guarantee a full protection to the side parts.
The eye safety of the operator is ensured by the
eye safety filter which is firmly installed into the slit
lamp. The eye safety filter is already included and
may not be removed at any time. If the CLASSIC
514 is used with the laser adaptor L-ORD (chapter
7), the eye safety filter is included. Therefore, no
further safety measures are necessary for the
operator.
A T T E N T I O N
Never look directly at the laser beam or the light reflected by the laser beam. This will
cause serious damage to your eye.
4.3 Electrical protection
Never remove the housing parts of the laser. Work on the device may only be carried out
by authorized A.R.C. Laser GmbH service personnel. The mains may only be replaced
by authorized service personnel.
The room in which the laser is operated should be kept dry. In the event that cleaning is
necessary, make sure that the floor is dry before starting up the device.
A T T E N T I O N
Never use the laser if you notice any visible defect on the device.
Never use the device if you notice any visible defect on the power plug, wires or supply
lines are exposed.
Содержание CLASSIC 514
Страница 53: ...User Manual CLASSIC 514 August 2023 File User Manual_CLASSIC 514_rev5 docx 53 Notes...
Страница 54: ...User Manual CLASSIC 514 August 2023 File User Manual_CLASSIC 514_rev5 docx 54 Notes...
Страница 55: ...User Manual CLASSIC 514 August 2023 File User Manual_CLASSIC 514_rev5 docx 55 Notes...
Страница 56: ...A R C Laser GmbH Bessemerstr 14 D 90411 N rnberg Tel 49 911 21779 0 Fax 49 911 21779 99 www arclaser de...