Place the L-ORD on the pin of the tonometer mounting base at the slit lamp, as shown in
the image below. Turn the lockwing screw thight to fix the L-ORD.
A T T E N T I O N
The laser can only be set into READY mode, if the L-ORD swung in correctly in the
working position. If the L-ORD is not correctly in working position an error message
occurs. Please check the position of the L-ORD in this case.
After the L-ORD is adapted to the slit lamp, the connection hose should be fixed with the
provided clamps at the chin rest.
L-ORD swung out
Digonosis position of the slit lamp
L-ORD swung in
Working position
– Laser radiation is possible