User Manual CLASSIC 514 

 

August 2023

 

 

 

File: User Manual_CLASSIC 514_rev5.docx 

10 

 

Set-up and installation 

3.1  Installation site 

Before  the  device  is  delivered,  it  must  be  ensured  that  the  laser  can  be  set  up  in  a 
suitable location. 

The  CLASSIC  514  should  be  operated  in  an  easily  accessible  place.  The  laser  should 
not  be  operated  near  a  heater,  since  the  device  only  works  when  the  ambient 
temperature  is  not  higher  than  35° C.  Higher  temperatures  can  cause  the  device  to 
switch off due to overheating. A place of installation in direct sunlight can cause an early 
shutdown and must be avoided in any case. 

If  the  temperature  is  too  low  (below  +18° C),  the  device  cannot  be  started  to  prevent 
condensation on the internal optics; this could result in permanent damage to the laser. 

 

The device should be set up so that the laser beam is not aimed directly at a door, 
window or reflective material. 

 

The device should be operated in a dust-free room. There should be no carpets on 
the floor or the walls. 

 

When the device is not in use, it should be covered to prevent contamination. For 
this purpose, a suitable dust cover is included. 

The wiring must be installed so that there are no tripping hazards or other hazards 

All  control  elements  must  also  be  free  and  easily  accessible.  The  air  humidity  is 
monitored internally and must be below 75 %. 

3.2  Room requirement 

The legislation imposes the following requirements to any room in which a class 4 laser 
according to IEC 60825-1 is operated. 

3.2.1  Marking access points 

All entrances must be clearly marked so that unintentional or unauthorized entry, which 
can lead to a hazard, is avoided. 

 

Laser warning signs (triangle with laser symbol) and the wavelength labeling must 
be attached to each access door. 

 

A warning lamp must be attached above each access door. This must always light 
up when the laser is in operation. 
Unintentional entry into the room without safety glasses is avoided. 

 

The laser safety goggles must be easily accessible at the entrance. 

3.2.2  Window shielding 

It  must  be  ensured  that  no  laser  radiation  can  escape  the  room.  In  particular,  the 
windows  must  be  secured  with  suitable  materials.  If  you  have  any  questions,  your 
contact person at A.R.C. Laser GmbH is available to assist you. 

 

 

Содержание CLASSIC 514

Страница 1: ...User Manual A R C Laser GmbH 0123 Bessemerstr 14 D 90411 N rnberg Tel 49 911 21779 0 Fax 49 911 21779 99 info arclaser de www arclaser de...

Страница 2: ...Bessemerstr 14 D 90411 N rnberg Tel 49 911 21779 0 Fax 49 911 21779 99 service arclaser de Copyright A R C Laser GmbH All rights reserved This document contains confidential and copyrighted informatio...

Страница 3: ...ective housing 15 4 10 Connectors and switches 15 4 11 External interlock switch 16 4 12 Safety Shutter Aiming beam 16 4 13 Manual reset 16 4 14 Reset by power failure 16 4 15 Lables and markings 17 4...

Страница 4: ...used with the L ORD 41 8 2 2 Classification 41 9 Maintenance 42 9 1 Introduction 42 9 2 Safety Check STK 42 9 3 Care by the user 42 9 4 Cleaning of Accesories 44 9 4 1 Foot switch 44 9 4 2 Laser safe...

Страница 5: ...If you have any further questions regarding the safety or the use of the device or regarding the laser and laser radiation please contact A R C Laser GmbH 1 1 Marking and symbols The symbol Attention...

Страница 6: ...n Infection inflammation iritis from iridotomy or trabeculoplasty posterior synechiae from iridotomy anterior synechiae from trabeculoplasty bleeding not persistent IOP increase after treatment macula...

Страница 7: ...on arises To improve this effect the laser is equipped with a resonator in our case the photon emitting semiconductor itself The plan parallel properties of the crystal with its high index of refracti...

Страница 8: ...750 hPa During storage a temperature range of 5 C to 40 C must be maintained The environment air must be dry and clean The relative humidity during transport and storage must be between 5 and 85 A T...

Страница 9: ...GmbH be sure to use the intended transport packaging Normally the CLASSIC is delivered together with the slit lamp For transportation the device is packed as follows The transport packaging is design...

Страница 10: ...n use it should be covered to prevent contamination For this purpose a suitable dust cover is included The wiring must be installed so that there are no tripping hazards or other hazards All control e...

Страница 11: ...matted In the area of laser use only matted non reflective and non flammable instruments and materials should be used 3 3 Electrical connection The laser is operated with a DC voltage of 24 V The pow...

Страница 12: ...should enclose applicable legal regulations and provisions with the device and inform the staff accordingly If the device is operated outside Germany provisions of the American National Standard Offic...

Страница 13: ...esulting from internal reflections on the glasses Therefore we encourage you to wear laser safety glasses which also guarantee a full protection to the side parts The eye safety of the operator is ens...

Страница 14: ...e irradiance is equal to the exposure limit value of the cornea of the eye The NOHD identifies the danger area within which there is a risk of damage to the health of the eye if you look directly into...

Страница 15: ...tances in concentrations whose placing on the market is prohibited according to the applicable requirements of Directive 2015 863 EU RoHS 3 4 9 Protective housing The CLASSIC 514 laser has a protectiv...

Страница 16: ...standard which can be replaced by a door interlock switch The laser switches off when the door is opened In addition the laser cannot be switched to the READY mode when the door is open An error messa...

Страница 17: ...gs in accordance with European and worldwide guidelines to prevent the laser user from being exposed to laser radiation by negligence The following labels can be found on the CLASSIC 514 Type plate 1...

Страница 18: ...80 hPa and 750 hPa The following environmental conditions must be met Ambient temperature 18 C to 35 C Relative humidity 75 4 17 Electromagnetic compatibility The CLASSIC 514 laser system meets the EM...

Страница 19: ...es from portable or mobile wireless communication devices please refer to Chap 18 3 ELECTROMAGNETIC PROTECTION II and 18 4 RECOMMENDED PROTECTION DISTANCES RFID systems shall not be used at separation...

Страница 20: ...instruction from an A R C Laser GmbH employee or by an authorized representative when installing the device During instruction the use of laser safety glasses and laser safety will be specifically add...

Страница 21: ...accessories transduce and services other than those which the A R C Laser GmbH has determined or provided may result in increased electromagnetic interference or reduced electromagnetic immunity of th...

Страница 22: ...CLASSIC 514 The laser radiation which is coupled in the fiber connector is re directed through the zoom optic into the beamcombiner For using the laser with the slit lamp the beamcombiner needs to be...

Страница 23: ...ems are duly packed by A R C Laser GmbH and supplied together with the corresponding accompanying documentation of the respective manufacturers This accompanying information has to be respected Before...

Страница 24: ...to malfunctions The laser is in the READY mode during treatment If you need to interrupt the laser must be returned to STANDBY mode The device must always be switched off when unattended to prevent o...

Страница 25: ...d herein may result in hazardous laser radiation exposure 1 Laser Stop 7 Adjustment of pulse length with plus minus 2 Laser READY LED 8 Adjustment of pulse pause with plus minus 3 Laser READY button 9...

Страница 26: ...Starting the laser The CLASSIC 514 is used together with a slit lamp The device is connected via a power supply to the lifting table To start the device and the slit lamp switch on the main switch und...

Страница 27: ...apter 6 4 1 to change the password To enter the password turn the rotating wheel to the right The selected number is confirmed by pressing the rotating wheel Afterwards a star in the upper code line m...

Страница 28: ...y pulse pause settings In the factory settings the right arrow keys can be used to change the frequency Hz up to 10 Hz Following values are possible 1 Hz to 5 Hz in 1 Hz steps 10 Hz and SP signle puls...

Страница 29: ...to 15 The device emits 5 pulses per 10 ms which means 25 pulses when chosing 50 ms 50 pulses when chosing 100 ms and so on The percentual values 5 10 15 correspondend to the share of the delivered po...

Страница 30: ...eel changes the respective value If an icon is darkened it is active and the values can be changed 1 Display brightness can be adjusted from 25 100 in 5 steps 2 Audio frequency can be adjustet from 0...

Страница 31: ...bol redirects you into another sub menu This menu contains information concerning your CLASSIC 514 Furthermore it informs about the last occurring error which you can communicate to the service person...

Страница 32: ...rt at a low level and then increase energy to the target during treatment Fix the patient s eye with a contact glass and focus the target beam into the eye to be treated Select the necessary beam inte...

Страница 33: ...will be pressed during treatment the same error message appears on the screen and the device cannot be used anymore A T T E N T I O N In general the Laser Stop should only be used in emergencies 6 9...

Страница 34: ...Audio pitch 0 100 Back in APL Mode Audio volume 10 100 Delete during entering password Aiming beam brightness 1 100 Accept during entering password Info menu Temperature too high Leave menu Temperatur...

Страница 35: ...Please use the packaging material of A R C Laser GmbH only to protect the device from damages during transport The L ORD should only be stored and transported in a temparture range between 10 C and 60...

Страница 36: ...e L ORD is used together with the CLASSIC 514 With regard to ambient conditions such as humidity and air pressure please refer to chapter 4 16 Due to the internal design of the L ORD a maximum tempera...

Страница 37: ...tand Make sure that the connection hose with the control cable of the L ORD is which is firmly attached to the connection hose is additionally connected to the red socket at the rear panel of the CLAS...

Страница 38: ...aser can only be set into READY mode if the L ORD swung in correctly in the working position If the L ORD is not correctly in working position an error message occurs Please check the position of the...

Страница 39: ...tion of the beam can be adjusted via this adjusting wheel Vertical Alignment The vertical position of the beam can be adjusted via this adjusting wheel Deflection mirror The deflection mirror is used...

Страница 40: ...has been developed according to the latest technical knowledge We have set the product life to 7 years In addition the availability of the psare parts is guaranteed by us within a period of 10 years A...

Страница 41: ...htness Power transmission coupled in a slit lamp with 62 5 m fiber 8 1 3 Electrical Connection Data Power supply connection values 100 240 V 50Hz 60Hz 1 06 0 45 A 8 1 4 Classification Laser class work...

Страница 42: ...be subjected to a safety check STK by trained personnel Scope of Safety Check STK Visual Check Laser marking laser class max Power wavelength Information signs warning signs properly and completely at...

Страница 43: ...s according to the manufacturer of the disinfectant the disinfectant should be suitable for wipe disinfection of surfaces the disinfectant should be based on alcohol and or quaternary compounds the di...

Страница 44: ...ssary clean the slit lamp mirror with a suitable lens brush After dedusting clean the mirror with Kodak lens cloths and a few drops of pure acetone Do not use the cloth dry otherwise the mirror will b...

Страница 45: ...tem check and have to be checked by the user Problem Proable error Troubleshooting No aiming beam visible Laser is in STANDBY mode Aiming beam is only visible when the laser is in READY mode Change to...

Страница 46: ...regulations must be respected for proper disposal Under no circumstances should the device be disposed of with domestic waste A R C Laser GmbH is happy to assist with the disposal of the system Costs...

Страница 47: ...s that are returned during the warranty period at the express request of A R C Laser GmbH must be confirmed in writing by A R C Laser GmbH Detailed packaging instructions and information on how to ret...

Страница 48: ...nity tests Compliance Electromagnetic environment guideline RF Emissions CISPR 11 150 kHz 30 MHz EN 55011 Group 1 Class B The laser uses RF energy exclusively for its internal function Hence RF emissi...

Страница 49: ...agnetic fields with energetic design frequencies 30 A m 50 Hz or 60 Hz IEC 61000 4 8 Magnetic fields at the grid frequency that are usuallty available in professional healthcare facilities may be used...

Страница 50: ...Power interruption 0 UT 250 300 cycles at 0 and 180 level IEC 61000 4 11 Only the voltage of a typical professional healthcare facility may be used It is recommended to use an uninterruptible power s...

Страница 51: ...d no closer than 30 cm 12 inches to any part of the CLASSIC 514 including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result The field strengt...

Страница 52: ...cation device as stated below Test frequency MHz Frequency band MHz Radio service Modulation Maximum performance W Distance m Immunity test level V m 385 380 to 390 TETRA 400 Pulse modulation 18Hz 1 8...

Страница 53: ...User Manual CLASSIC 514 August 2023 File User Manual_CLASSIC 514_rev5 docx 53 Notes...

Страница 54: ...User Manual CLASSIC 514 August 2023 File User Manual_CLASSIC 514_rev5 docx 54 Notes...

Страница 55: ...User Manual CLASSIC 514 August 2023 File User Manual_CLASSIC 514_rev5 docx 55 Notes...

Страница 56: ...A R C Laser GmbH Bessemerstr 14 D 90411 N rnberg Tel 49 911 21779 0 Fax 49 911 21779 99 www arclaser de...