9  Service 

9.1  Introduction 

The device was designed, developed and tested according to research based on state-of-
the-art  technology.  As  a  rule,  similar  devices  have  a  service  life  exceeding  5  years.  This 
product  service  life  is  defined  to  be  10  years  in  order  to  guarantee  the  availability  of  its 
spare  parts  during  this  period  of  time.  However,  to  make  sure  the  product  is  in  good 
working order, the device can display its internals working status through the touch screen 
display.   

In  the  rare  sit

uations when the device does not work, you can use the “Troubleshooting” 

section  included  in  this  manual,  to  isolate  the  problem.  The  solutions  given  therein  will 
most  likely  allow  you  to  solve  the  problem  yourself.  However  if  the  problem  cannot  be 
solved, please contact your local A.R.C. Laser authorized dealer for technical support. 

A T T E N T I O N ! 

Do  not  attempt  to  perform  any  type  of  service  or  maintenance  work  to  the  device.  Any 
calibrations  or  adjustments  that  require  to  open  the  protective  housing,  should  only  be 
carried out by a service technician trained by A.R.C. Laser GmbH. This includes also any 
type of optics cleaning within the laser system.

 

This includes also any type of optics cleaning within the laser system. 

9.2  Care and Maintenance (by the user) 

9.2.1  Cleaning the exterior surfaces 

Wipe  the  outer  surface  with  a  clean,  lint-free cloth. If  necessary,  the  cloth  can  be  slightly 
moistened  with  a  non-corrosive  cleaning  solution  or  isopropyl  alcohol.  Rub  with  a  clean, 
lint-free cloth or allow air to dry. Never spray the C-Pro or apply detergents. 

Caution: The micromanipulator must never be immersed in cleaning liquid, nor should it be 
placed  in  the  autoclave.  This  will  cause  irreversible  damage  to  the  optical  components. 
The  micromanipulator  cannot  be  sterilized.  To  ensure  the  sterility  of  the  surgical  field, 
sterile coatings must be used. 

9.2.2  Cleaning the deflection mirror 

The  deflecting  mirror  must  be  cleaned  in  case  of  contamination  or  in  case  of  power 
malfunctions. Contamination on the deflecting mirror creates spots on the mirror surface. 

Only two cleaning methods should be performed: 

1. Compressed air cleaning

Compressed air can be used to remove particles from the surface 

2. Moisture cleaning

Cleaning  with  methanol  can  be  carried  out  if  complete  cleaning  is  necessary.  The 
procedure is as follows. 

Содержание C-Pro

Страница 1: ...User Manual C Pro Micromanipulator Medical Device A R C LASER GmbH Bessemerstr 14 90411 N rnberg Germany phone 49 911 21779 0 fax 49 911 21779 99 info arclaser de...

Страница 2: ...t A R C LASER GmbH Bessemerstr 14 D 90411 N rnberg Germany TEL 49 911 21779 0 FAX 49 911 21779 99 Revision Date Nov 22 2017 File Manual_C PRO_10600nm_rev3 docx Revisions Date 12 10 2020 Author Christi...

Страница 3: ...or 12 4 4 Connecting the C Pro to the Surgical Microscope 12 4 4 1 Microscope compatibility 12 4 4 2 Connecting the C Pro 13 4 5 Connecting the Micromanipulator to the Laser 14 4 5 1 Connecting the ar...

Страница 4: ...trolling the laser beam 29 7 4 Dismantling 29 7 5 Quick guide 30 7 6 C Pro performance test after installation 30 8 TECHNICAL DATA 31 8 1 Dimensions 31 8 2 General 31 9 SERVICE 32 9 1 Introduction 32...

Страница 5: ...ntire user manual without an explicit written consent by A R C Laser GmbH Approved copy or parts extracted from this user manual as licensed by A R C Laser GmbH shall include references to the origina...

Страница 6: ...laser beam The main fields of application are indications in the field of ENT and gynecology CAUTION The device may only be operated by specially trained personnel who have knowledge of the medical e...

Страница 7: ...the beam in the field of vision Fig 1 Beam path at the micromanipulator 2 2 Mirror and Lens Based Systems Micromanipulators are divided into mirror based and lens based systems depending on their stru...

Страница 8: ...e C Pro instead of lenses as is common with many other manufacturers If the operating field is very weakly illuminated two target beams may be seen due to the light refraction at the deflecting mirror...

Страница 9: ...or the connection and the application of the micromanipulator see figure 4 Note The C Pro and all accessories should always be stowed in the appropriate recesses in the case after use The joystick and...

Страница 10: ...ease inform A R C Laser GmbH as soon as possible in case parts are missing or damaged 3 3 Storage Conditions Transport and storage conditions Temperature 10 C to 60 C Relative humidity 5 85 The C Pro...

Страница 11: ...o light up when the laser is in operation This ensures warning to prevent accidental entry to the room without protective goggles Please store the laser protective goggles at the entrance of the room...

Страница 12: ...t the top and the underside of the deflection mirror for burn in and contamination Hold the micromanipulator in a firm grip 4 Note Never touch the coated side of the deflection mirror 5 If cleaning is...

Страница 13: ...see figure 5 3 Push The dovetail adapter attached to the micromanipulator into the adapter plate so that the lens of the microscope is located centrally above the housing of the micromanipulator As a...

Страница 14: ...e thread adapter Make sure the articulated arm is in the correct position see figure 7 8 The articulated arm is taken out of the holder by means of slight pressure of the two fingers so that the conne...

Страница 15: ...trated operating area The laser must be positioned in such a way that the articulated arm is within the operating range shown in figure 10 4 6 2 Angular position of the articulated arm The two arms of...

Страница 16: ...line with the angular range mark red green red label on the first leg see figure 11 right After the laser has been placed in the positions described in 4 5 1 and 4 5 2 the laser and the operating mic...

Страница 17: ...r In addition you can refer to the regulations of the American National Standard Agency ANSI Z136 3 1996 American National Standard for the Safe Use of Lasers in Health Care Facilities and ANSI Z136 1...

Страница 18: ...protected from catching fire This can be achieved by the use of fire resistant instruments venturi ventilation humidification and non burning gas mixtures The anesthesia staff should use fire resistan...

Страница 19: ...address reorders to A R C Laser GmbH indicating the wavelength or the item number of the goggle Should you require any additional information on protective goggles please contact your local A R C Las...

Страница 20: ...materials as this may cause a fire The pedal switch controlling the laser Pulse should never be outside the range of the operator holding the laser hand piece fiber It is prohibited that any person ot...

Страница 21: ...company operates worldwide and considers the protection of the environment and natural resources as a corporate obligation Based on individual tests we can for the products of ARC Laser GmbH confirm t...

Страница 22: ...been tested for operation in heights exceeding 2 000 m above sea level The following ambient conditions must be met Operating conditions Temperature 15 C to 35 C Relative humidity 75 not condensing A...

Страница 23: ...n training all trained persons are mentioned in the device book wherein a person as responsible for the security is registered This person is subsequently entitled to instruct other people on the unit...

Страница 24: ...s equipped with the following components see figure 13 Fig 13 Components on the C Pro connection for articulated arm adapter positioning ring operating distance dovetail focusing defocussing lever def...

Страница 25: ...na Vario 1 x ME02157 thread adapter for articulated arm 1 x ME12033 hand rest long 1 x VP12000 transportation case C Pro 1 x ZU01084 cleaning set selected when ordering For information on accessories...

Страница 26: ...the beam diameter is set the lower the energy density for the same laser parameters The power density is defined by the following formula with The unit of is In the table below the energy densities a...

Страница 27: ...cused before the target beam can focused see figure 14 3 Get the laser ready Press the grey button standby in the lower right corner of the display It will change into yellow button ready 4 The desire...

Страница 28: ...this lever position for tissue cutting Moving the lever in the direction D defocused will cause an increase in the beam diameter A defocused beam is used for coagulation and vaporization of tissue In...

Страница 29: ...ted by turning the cap Fig 17 C Pro from the surgeon s perspective left focus defocus center joystick right hand rest The size of the operating area depends on the operating distance Minimum 50 x 30 m...

Страница 30: ...be adjusted as required 7 6 C Pro performance test after installation To ensure that the C Pro is working well please perform the following test routine after installation 1 Put a wooden spatula in f...

Страница 31: ...dth x height Weight 570 g 8 2 General Designation Micromanipulator Type C Pro Transmission 70 5 Wavelength CO2 laser 10600 nm Maximum power CO2 laser 30W CW 10W CW Super Puls Operating distance 200 50...

Страница 32: ...trained by A R C Laser GmbH This includes also any type of optics cleaning within the laser system This includes also any type of optics cleaning within the laser system 9 2 Care and Maintenance by th...

Страница 33: ...ning set ZU01084 Pre cleaning For cleaning the mirror please turn the micromanipulator over so that the underside of the deflecting mirror points upwards Fig 20 Position of the C Pro at the beginning...

Страница 34: ...eflecting mirror by means of lens cleaning tissue 2 The cleaning tissue should overlap on the side facing the observer so that the observer can hold the tissue with the tweezers without touching the m...

Страница 35: ...he lens cleaning tissue may only be used once a fresh tissue should be used for each additional cleaning operation The steps described should be repeated as often as necessary until the mirror surface...

Страница 36: ...ination of the operating area Explanation see chapter 7 1 No target beam Switch on the laser Activate the target beam Check the operating distance of the adjusting ring Micromanipulator is loose Check...

Страница 37: ...sing of the device shall be passed on to A R C Laser GmbH within 24 hours Parts returned during the warranty period upon the explicit request of A R C Laser GmbH are subject to the written confirmatio...

Страница 38: ...Manual C PRO 38 Notes...

Страница 39: ...Manual C PRO 39 Notes...

Страница 40: ...Manual C PRO 40...

Отзывы: