5.10 Operating Conditions:
The Micromanipulator C-Pro is not suitable for use with combustible gas mixtures of all
kind.
The device has not been tested for operation in heights exceeding 2,000 m above sea
level.
The following ambient conditions must be met:
Operating conditions
Temperature
+15°C to +35°C
Relative humidity
< 75% (not condensing)
A T T E N T I O N !
The micromanipulator must acclimatize for at least 4 hours at room temperature after
transport. This prevents condensation on the optics.