5.4 Explosion and Fire Hazard
A T T E N T I O N !
Never work with the laser in the vicinity of easily flammable anesthetics such as oxygen
and oxygen-enriched atmosphere (special mechanical ventilation!) easily flammable
solutions or material. In particular, please remove combustible plastic or paper elements
from and around the operating area of the laser. Focusing the green or IR laser beam on
flammable materials may ignite these and cause a fire or explosion!
When working with the laser, make sure the laser is switched from READY mode to
STAND-BY mode, in case the treatment has to be interrupted. This will assure that no
laser radiation will be emitted due to unintentional stepping on the pedal switch.
5.5 Protection against Undesired Radiation
Never grip the hand piece or fiber tip with their exit pointing or touching your exposed skin
as this may cause burns to the skin. Do not focus the laser beam on easily flammable
materials as this may cause a fire.
The pedal switch controlling the laser Pulse should never be outside the range of the
operator holding the laser hand piece/fiber. It is prohibited that any person other than the
operator controls the pedal switch.
In operating rooms in which several pedal switches are available, it is particularly important
to make sure that the laser pedal switch is in within the operator vicinity.
A T T E N T I O N !
During a laser treatment, the system is in "Ready" mode. Should the operator need to
pause the treatment for any reason; the laser must be switched back to "Stand-by" mode.
In case of pauses exceeding 1 minute, please press the "Ready" key.
The device must be turned off when left without supervision, to prevent usage by
unauthorized individuals.
5.6 NOHD Safety Distance
The values listed below are based on:
BGI 832 -
Operation on Laser Devices
Application of safety regulations
„Laser Radiation“ BGV B2 on new laser classes and MZB-values according to DIN EN 60
825-1 (VDE 0837-1): 2001-11
Trade association of Precision Engineering and Electrical Engineering BGFE dated April
2003(ZH1/405)
Wavelength
λ:
10600 nm
NOHD:
1.4 m