Set-points
Temperature:
55.0 °C
Preheat:
0 min
Air removal:
16.0 kPa
Chamber test rate:
0.3 kPa / min
Conditioning:
60% 30 min
EO injection:
16.0 kPa
EO exposure:
1:00 hh:mm
EO removal:
16.0 kPa
Flushing (count):
5
Aeration
Programmed
12:00 hh:mm
Figure 83
Example Cycle Report ‑ Set‑points Option
Cycle complete
No errors
No cautions
End time:
04/20/2015 21:03:12
59%
55.0° C
1:00 hh:mm
12:00 hh:mm
15:30 hh:mm
RH end of
conditioning:
Temp end of
conditioning:
Actual expose time:
Actual aeration time:
Total cycle time:
Reviewed by & date:
Figure 84
Example Cycle Report – Footert
Reference Section 1113 for a obtaining a detailed
report of set points for custom cycles
The definitions for the column headings in the Table cycle report (Figure 82) are as follows:
•
Tc (C)
is the chamber temperature (°C) read by the control sensor
•
Tm (C)
is the chamber temperature (°C) read by monitor sensor
•
Tld (C)
is the load zone temperature (°C) of the chamber heating blankets
•
Tmid (C)
is the middle load zone temperature (°C) of the chamber heating blankets
•
Tul (C)
is unload zone temperature (°C) of the chamber heating blankets
•
Pc (mBar)
is the chamber pressure (mBar) read by the control sensor
•
Pm (mBar)
is the chamber pressure (mBar) read by monitor sensor
•
HS (C)
is heatsink (vaporizer) temperature (°C)
•
RHc (%)
is the chamber percent relative humidity (%RH) read by the control sensor
•
RHm (%)
is the chamber percent relative humidity (%RH) read by monitor sensor