50
3M
™
Steri‑Vac
™
Sterilizer / Aerator GSX Series
– Operator’s Manual
11.10. Removing Custom Cycles from the 3M™ Steri‑Vac™ Sterilizer/Aerator GSX Series
1�
To remove a custom cycle on the 3M™ Steri‑Vac™ Sterilizer/Aerator GSX Series, the cycle can be deleted with the
Delete
cycle
button and/or the cycle can be copied to a USB drive To
copy to a USB drive, place a USB drive into one of the GSX Series sterilizer USB ports
2�
Press
Browse
to search for valid Folders on the USB drive Choose the desired USB Folder by pressing the name, then
OK
3�
The GSX Series sterilizer will automatically create a valid USB folder on the USB drive to store the exported cycle(s) (eg “\ USBDisk \ Custom Cycles GS5X \ EA343434 Export cycles”)
(Figure 61)
4�
Select single file for export from the right window and click (“<”) To export all files, click (“<<”) The selected files will move to the left window (Figure 61)
5�
A green check mark next to the file name indicates the custom cycles are ready to be exported (Figure 61) Select
Export
cycles
6�
The sterilizer will ask for confirmation of the Folder designation before exporting
7�
Press
Return
to go back to the main screen
Sterilizer
22.2
˚
C
04/20/2015
14:07:49
98.0
kPa
Scope 395
Stent 126
Import/export cycles
USB folder
...\Custom Cycles GS5X\EA343434 Export cycles
Cycles on sterilizer
\USBDisk\Custom Cycles GS5X
...
Export cycles
Return
Delete cycle
Import cycles
Cancel
Scope 395
Figure 61
USB Folder for Exporting