26
3M
™
Steri‑Vac
™
Sterilizer / Aerator GSX Series
– Operator’s Manual
The minimum default aeration time on the 3M™ Steri‑Vac™ Sterilizer/Aerator GSX Series can be changed for the preprogrammed GSX 38C and GSX 55C cycles Press the
minimum aeration
time and an iWheel will appear to adjust the minimum aeration default time Figure 23 illustrates changing the minimum aeration default time
Minimum aeration times cannot be changed in this view for custom cycles Custom cycle minimum aeration times must be changed in the 3M™ Cycle Programmer and imported back to the
GSX Series sterilizer
Sterilizer
22.2
˚
C
04/20/2015
12:38:29
98.0
kPa
Select cycle temperature:
6
5
4
3
2
1
7
Return
Apply changes
Cycle access level
Touch check box to select cycles listedon start screen.
Touch row to change values.
Operator
Cycle
2.0 hrs
2.0 hrs
Minimum aeration
Setup
Preferences Cycle Setup Date-Time Network Site
55 ˚C
Minimum aeration
10
11
12
00
hh:mm
:
30
13
14
OK
OK
Cancel
Cancel
GSX 38C
GSX 55C
Figure 23
Changing the Minimum Aeration
Default Time
The
Site Setup>Setup tab
(Figure 24) provides the specific configuration and maintenance details for your site; this is established by authorized 3M service personnel during installation
Sterilizer
22.2
˚
C
04/20/2015
12:39:27
98.0
kPa
Select cycle temperature:
6
5
4
3
2
1
7
55 °C
Return
Apply changes
Ventilation hood
High elevation
Abator always on
One door
Abator mode
Lock out immediate
Abator setting
Next maintenance
10-2015
Reset
Reset
Barcode override count
0
System cycle count
38
Last maintenance
04/20/2015
Maintenance interval
6 months
Setup
Preferences Cycle Setup Date-Time Network Site
Ventilation hood
High elevation
One door
1 to 1
Figure 24
Site Setup>Setup Tab ‑ Options