64
65
13.2. Display Screen Indications
Figure 79 illustrates the important information presented on the display screen while the sterilization cycle is running Table 5 explains each indication
6
Sterilizer
55.0
Menu
Menu
˚
C
04/20/2015
01:41:59
Cycle:
Count:
33
GSX 55C
Show chart
98.0
kPa
Stop
Stage:
Flushing
C604 - Power interruption
Details...
04/20/2015 06:50:57
EO
E109 -
Details...
04/20/2015 01:38:08
Open door
Open unload
door
0:00
Cycle time remaining (hh:mm):
1:00
Gas expose time elapsed (hh:mm):
12:00
Aeration time elapsed (hh:mm):
Cycle failure
Error level recovery
Cycle complete
RH
50
%
8
8
9
1
2
3
4
5
7
10
11
12
13
14
15
16
17
18
Figure 79
Display Screen Indications
Indication Number
Indication Description
1
EO Icon.
Indication ethylene oxide (EO) is in the chamber
2
Shows current chamber temperature
3
Shows current chamber pressure
4
Shows current chamber percent relative humidity (%RH) %RH is only measured and displayed during the Conditioning stage
5
Indication that Maintenance is due and Service should be notified
6
An orange, downward arrow symbol will appear on the display screen task bar to remind the user that a pending software update is available for installation
7
Shows current date and time
8
Error Message.
Indicates a fault has occurred See Chapter 16: Cautions, Error Messages and Troubleshooting for specific errors and corrections
9
Indication that a
used
3M™ Steri‑Gas™ EO Gas Cartridge is in the chamber
10
Indication that an
unused
3M™ Steri‑Gas™ EO Gas Cartridge is in the chamber
11
Counter.
This is the sterilizing cycle counter that indicates the cumulative number of all cycles started, including those cycles in which a fault occurred
12
Cycle stage icons.
While a sterilization cycle is running, the icons on the bottom of the display screen indicate the current stage of the cycle (indicated by a white
border around the icon) Each icon and the corresponding sterilization cycle stage are described in the section Overview of the GSX Series EO Sterilization Cycle
13
Menu
button is used to access GSX Series sterilizer options (eg cycle reports, setup variables)
14
Open Door
button (the word “load “is only used for double door or pass through sterilizer models) During the sterilization process, the Operator cannot open the
door; therefore, the button is colored gray
15
Open Unload Door
button is used on double door pass through GSX Series sterilizer models During the sterilization process, the Operator cannot open the door;
therefore, the button is colored grey
16
Show Chart
button displays a running graph of the current cycle in process
17
Stop
button cancels the cycle in progress There are times during the sterilization process when the Operator cannot cancel the GSX Series sterilizer; therefore, the
button is colored grey
18
Cycle Complete
signifies the Operator programmed cycle is successfully complete
Table 5. Explanations of Display Screen Indications