50
51
11.11. Cycle Reports
11111 Custom Cycle Set Points
The 3M™ Steri‑Vac™ Sterilizer/Aerator GSX Series allows users to view and print parameter values for custom cycles installed on the GSX Series sterilizer The report will display programmed
custom cycle set‑points for each parameter
1�
To view or print custom cycle set points go to
Menu>Reports>Custom cycle set points
2�
Select the desired cycle and then
View
or
Figure 62 is an example custom cycle set point report
3�
Press
Cancel
to return to the main screen
Custom cycle report
EA343434 04/20/2015 15:13:27
Software version 1.0.15
--------------------------------------
Cycle: 1 Test cycle
Stage: Preheat
Cycle temperature = 60 °C(140 °F)
Timeout = 60 min
Hold time = 0 min
Stage: Air removal
Heatsink temperature = 95 °C(203 °F)
Chamber pressure = 160 mbar
Duty cycle = 100 %
Stage: Chamber test
Chamber test time = 5 min
Stage: Conditioning
Relative humidity = 40 %
Humidify timeout = 50 min
Hold time = 0 min
RH injection rate = Fast
RH threshold lower = 10 %
RH threshold upper = 90 %
Stage EO injection
Simulate EO = Yes
Inject pressure = 160 mbar
Minimum pressure increase = 200 mbar
RH threshold lower = 20 %
RH threshold upper = 60 %
Stage: EO exposure
Exposure time = 1 min
Stage: EO removal
Stage: Flushing
Number of purge cycles = 5
Vent duty cycle = 100 %
Dwell time top = 4 min
Pump down pressure = 300 mbar
Dwell time bottom = 0 min
Purge aeration time = 90 min
Mandatory locked aeration = 180 min
Stage: Aeration
Aeration temperature = 34 °C(93 °F)
Minimum aeration time = 0.5 hr
Figure 62
Example Custom Cycle Set Point Report