7 Specifications
7.1. 3M™ Steri‑Vac™ Sterilizer/Aerator GSX Series Structural Specifications
GS5X
GS8X
Operational Weight
127 kg (281 lbs) single door
132 kg (290 lbs) double door
261 kg (576 lbs) single door
269 kg (593 lbs) double door
Exterior Dimensions
H 709 cm x W 762 cm x D 950 cm
H 279 in x W 300 in x D 374 in
H 1798 cm x W 940 cm x D 1090 cm
H 708 in x W 370 in x D 429 in
Chamber Internal Volume
136 L (48 cubic feet)
224 L (79 cubic feet)
Chamber Internal Dimensions
H 380 cm x W 430 cm x D 830 cm
H 150 in x W 170 in x D 325 in
H 460 cm x W 510 cm x D 970 cm
H 180 in x W 200 in x D 380 in
7.2. Sound Power Levels Specifications
GS5X
GS8X
A‑weighted
< 85 dBa
< 85 dBa
7.3. Power Specifications
WARNING: To reduce the risk of shock due to hazardous voltage,
the customer must provide a properly grounded outlet (an earth ground) for installation as described in the installation requirements section of this manual
CAUTION: To reduce patient risks associated with exposure to potentially non‑sterile devices or inadequate sterilization procedures,
do not place any device emitting strong electronic magnetic fields (EMFs) near the sterilizer
Electrical Power
Operating Condition
Units
Voltage Range
200 ‑ 240
VAC
Frequency
50/60
Hertz
Phase
Single
Not Applicable
GS5X Current
7
Amps
GS8X Current
12
Amps