SEITE 4 - 148 x 210 mm - 44000118156/02 - SCHWARZ - 20-115 (kn)
4
Light emission area: 60-65 mm
2
(optically active)
Intermittent operation: : The device has been desi-
gned solely for short-term
operation.
Typical operating time at
room temperature (23° C):
7 min, at 40° C ambient tem-
perature: 1 min on, 15 min
off (cooling-off period)
Total exposure time
with new, fully
charged battery:
typically 120 min
Dimensions: Diameter: 28 mm
Length: 270 mm
Weight: 250 g (incl. light guide)
Charger and Handpiece
Time to charge
empty battery: approx. 1.5 h
Operating temperature: 10° C up to 40° C /
59° F up to 104° F
Relative humidity: 30% up to 75%
Atmospheric pressure: 700 hPa up to 1060 hPa
Total height with handpiece
inserted in the charger: 180 mm
Year of manufacture: see rating plate
Transport and Storage Conditions:
Ambient temperature: -20° C up to +40° C /
-4° F up to +104° F
Relative humidity: 30% up to 75%
Atmospheric pressure: 700 hPa up to 1060 hPa
Subject to technical modification without prior notice.
+40°C
+104°F
–20°C
–4°F
max.
75%
30%
hPa
1060
700
Installation of the Unit
Factory Settings
The factory settings of the unit are as follows:
• 10 sec exposure time
Initial Steps
Charger
E
Please ensure first that the voltage stated on the
rating plate corresponds to the existing main supply
voltage. The rating plate is attached to the bottom of
the unit.
E
Place the charger on a level surface.
E
Connect the power cable of the charger to the power
supply.
- The green LED on the charger showing that the
unit is ready for operation comes on. This shows
that the unit is ready for operation; please refer
to the section “Operating Status Display of the
Charger.”
Light Guide/Handpiece
E
Never place the handpiece in the charger without
the battery inserted in the handpiece!
E
Place the glare shield on the front of the unit.
E
Autoclave the light guide prior to first use.
E
Then attach the light guide firmly to the handpiece.
Inserting the Battery
E
Remove the protective cap from the battery and
keep with the unit packaging.
E
Slowly insert the battery with the thread side towards
the metal housing into the handpiece until it stops.
Screw the battery clockwise by hand until the seal is
pressed firmly against the metal housing. The unit
will not function properly if the battery is not screwed
in completely!
Содержание Elipar DeepCure-S
Страница 2: ...U2 148 x 210 mm 44000118156 02 SCHWARZ 20 155 kn ...
Страница 16: ...SEITE 14 148 x 210 mm 44000118156 02 SCHWARZ 20 115 kn 14 ...
Страница 48: ...SEITE 46 148 x 210 mm 44000118156 02 SCHWARZ 20 115 kn 46 ...
Страница 62: ...SEITE 60 148 x 210 mm 44000118156 02 SCHWARZ 20 115 kn 60 ...
Страница 76: ...SEITE 74 148 x 210 mm 44000118156 02 SCHWARZ 20 115 kn 74 ...
Страница 118: ...SEITE 116 148 x 210 mm 44000118156 02 SCHWARZ 20 115 kn 116 ...
Страница 132: ...SEITE 130 148 x 210 mm 44000118156 02 SCHWARZ 20 115 kn 130 ...
Страница 146: ...SEITE 144 148 x 210 mm 44000118156 02 SCHWARZ 20 115 kn 144 ...
Страница 174: ...SEITE 172 148 x 210 mm 44000118156 02 SCHWARZ 20 115 kn 172 ...
Страница 188: ...SEITE 186 148 x 210 mm 44000118156 02 SCHWARZ 20 115 kn 186 ...
Страница 202: ...SEITE 200 148 x 210 mm 44000118156 02 SCHWARZ 20 115 kn 200 ...
Страница 218: ...SEITE 216 148 x 210 mm 44000118156 02 SCHWARZ 20 115 kn 216 ...
Страница 222: ...SEITE 220 148 x 210 mm 44000118156 02 SCHWARZ 20 115 kn 220 ...
Страница 223: ...U3 148 x 210 mm 44000118156 02 SCHWARZ 20 155 kn ...