INTRODUCTION

Congratulations on the purchase of your new E-A-RTONE™ 3A or 5A Insert Earphones.  Whether 

you are performing comprehensive audiological diagnostic assessment, testing related to the fi tting 

or evaluation of amplifi cation, audiological research, or hearing screening in an educational or 

industrial setting, E-A-RTONE™ Insert Earphones provide many advantages that can enhance test 

reliability and minimize or eliminate problems typically encountered with supra-aural earphones.  

Because the test subject’s external canal is sealed with an E-A-RLINK™ foam eartip, reliable 

threshold measurement can be accomplished in environments previously unsuitable for hearing 

testing. The E-A-RTONE™ Insert Earphone’s direct coupling to the ear canal also provides greater 

sound separation between ears across the test frequency range. That means more instances where 

the need for masking is eliminated, or the level necessary for effective masking is lower.  Average 

testing time will decrease without sacrifi cing reliability.  Your preferred method of clinical masking 

does not need to be altered; simply substitute the greater inter-aural attenuation values cited in the 

Obtained Inter-aural Attenuation table on page 2. of this booklet.

E-A-RLINK™ foam eartips are available in three sizes; a standard (3A) size that will comfortably fi t 

most adult ear canals, a pediatric (3B) size for small ear canals, and a large (3C) size for very large 

or surgically altered ear canals.  Two additional sizes of eartips are available as “Infant Eartips” in 

3.5mm and 4.0mm versions to accommodate very small ear canals.  Disposable eartips eliminate 

the need for periodic cleaning of a headband and earphone cushions, and they facilitate infection 

control procedures.

The E-A-RTONE™ 3A and 5A Insert Earphones are available with two levels of impedance        

(10

Ω

 and 50

Ω

) to match the output of most commercially available audiometers. Your distributor 

or a 3M Auditory Systems Customer Service Representative can help you determine which version 

is appropriate for your equipment.  

Unless your audiometer was purchased with an E-A-RTONE™ 3A or 5A Insert Earphone             

pre-calibrated by the audiometric equipment manufacturer, the Reference Equivalent Threshold 

Sound Pressure Level (RETSPL) values and procedure cited in the ANSI S3.6-2004 Standard 

must be used to calibrate your equipment before initial use. Re-calibration should subsequently 

be performed as specifi ed by the audiometer manufacturer; typically on an annual schedule.   

Measured correction factors can be applied where it is necessary to alternate between an insert 

earphone and a supra-aural earphone with audiometers that do not provide a dual calibration option.

E-A-RLINK™ foam eartips are designed for single-use and should be replaced with new eartips 

for each test session performed.  Re-use of these disposable eartips could result in inadvertent      

intra- and inter-subject cross-infection via contact transmission, or “biological reactivity” problems 

with the presence of residual chemicals when antimicrobial agents are used in an attempt to clean 

and disinfect previously used foam eartips.  Do not reuse E-A-RLINK™ foam eartips.

This instruction booklet provides information that will allow you to perform hearing testing with 

your audiometer and E-A-RTONE™ Insert Earphones with the confi dence that test results will be 

comparable to, or in certain instances more reliable than, measurements performed with supra-aural 

earphones.  Uncertainty regarding the use and calibration of E-A-RTONE™ Insert Earphones can 

most often be resolved by reviewing this booklet and the American National Standard Specifi cation 

for Audiometers (currently, ANSI S3.6-2004). For information regarding replacement items, please 

contact your distributor or a 3M Auditory Systems Customer Service Representative.   

 WARNINGS

•  Reliable test results can be obtained with this earphone only when the audiometric equipment 

that it will be used with is calibrated to the device, before initial use, and periodically thereafter 

as specifi ed by the audiometric equipment manufacturer, or in accordance with applicable 

national or international standards.

•  The use of accessories and replacement parts, other than those supplied by the manufacturer 

of the E-A-RTONE™ Insert Earphone, or its authorized distributors, could result in invalid test 

results. 

•  E-A-RTONE™ 3A Insert Earphone front tubes must be replaced periodically, typically on an 

annual basis or whenever the tube(s) become brittle or perforated.  Do not cut the sound tube to 

forestall replacement.  A change of 10mm in the length of the sound tube will result in a change 

of the frequency response by as much as 0.5 dB at some frequencies.

•  E-A-RTONE™ 3A and 5A Insert Earphones are available in two different impedance 

versions (10 ohm and 50 ohm) to accommodate most commercially available audiometric 

equipment. Before initial calibration and use, confi rm that the output impedance requirements 

of the audiometric equipment and the E-A-RTONE™ Insert Earphone impedance                              

(as identifi ed in the Serial number) are the same.

•  E-A-RTONE™ 3A and 5A Insert Earphone replacement cables are not interchangeable 

between the two models, and in the case of the E-A-RTONE™ 5A Insert Earphone, cable 

replacement must be both model and impedance specifi c (5A 10 ohm or 5A 50 ohm).

•  E-A-RLINK™  foam eartips are designed for single use and should be replaced with new 

eartips for each test performed.  Re-use of these eartips could result in inadvertent intra- and 

inter- subject cross-infection via contact transmission, or “biological reactivity” problems with 

the presence of residual chemicals when antimicrobial agents are used in an attempt to clean 

and disinfect previously used foam eartips. Do not reuse E-A-RLINK™ foam eartips.

1. PRELIMINARY PROCEDURES

Examine each ear canal for obstruction(s), and visually determine the appropriate size of                  

E-A-RLINK™ foam eartip needed to fi t the test subject’s ear canals.  An airtight seal is necessary, 

and some test subjects may require different size eartips for the right and left ear. The standard 

size E-A-RLINK™ 3A eartip fi ts most adult ear canals, and the smaller diameter, beige color foam       

E-A-RLINK™ 3B eartips are available if the standard size eartip is too large for full insertion into 

small ear canals.  Larger diameter E-A-RLINK 3C eartips are also available to obtain a good seal 

for test subjects with wide or surgically altered ear canals.  A good seal and the proper insertion 

depth will provide the greatest ambient and inter-aural attenuation.  Deep insertion may be diffi cult 

to achieve in extremely small ear canals, even with the small 3B size eartip. In these instances, 

a shallow E-A-RLINK™ 3B eartip placement, or the use of the smaller 3.5 mm or 4.0 mm “Infant 

Eartips” should be employed.  Once the appropriate eartip size has been selected;

•  Secure the section of black tubing extending from the eartip onto the plastic sound tube 

nipple of the E-A-RTONE™  3A Insert Earphone’s front tube (Figure 1.) or,  directly  onto the           

E-A-RTONE™  5A Insert Earphone’s stainless steel case nipple (Figure 2.).

•  For the E-A-RTONE 3A Insert Earphone™, slowly roll the E-A-RLINK™ eartip into the 

smallest diameter possible, and while holding the black tubing where it connects with the tube 

nipple, insert the eartip deeply into the ear canal. The procedure is essentially the same for                

the E-A-RTONE™ 5A Insert Earphone except that once the eartip is compressed, instead 

of holding the black tubing, grasping the transducer case between the thumb and forefi nger 

will facilitate the correct insertion. The correct (“Full”) insertion depth is obtained when the 

outward-facing edge of the E-A-RLINK™ eartip is 2-3 mm inside the entrance of the ear canal 

as illustrated in Figure 3. 

•  Hold the E-A-RLINK™ eartip in place in the ear canal until it has expanded.
•  After the test is completed, remove the eartips and replace them with a new pair for the next 

test subject.

-10

0

10
20
30
40
50
60
70
80
90

100

110

125 250  500 1000 2000 4000 8000

Hz

dB

Obtained Interaural Attenuation

Full E-A-RLINK insertion in both ears
Shallow insertion in dead ear only
Shallow insertion in both ears

Table 1.

Full insertion depth, as described above, and illustrated in Figure 3., is necessary to achieve 

maximum ambient noise attenuation and inter-aural attenuation. Shallow insertion, as illustrated 

in Figure 4. may be the only option for certain ear canals, but more commonly refl ects incomplete 

compression of the eartip before insertion.  Table 1. below illustrates inter-aural attenuation as a 

function of the eartip insertion depth.  The described inter-aural attenuation is obtained for each ear 

in which the E-A-RLINK™ eartip is fully inserted, regardless of whether it is the “sending” ear or 

“receiving” ear.

2. BACKGROUND NOISE REDUCTION

A major benefi t of the E-A-RTONE™ Insert Earphone is the reduction of background noise that 

might otherwise mask air or bone conduction test signals and infl uence threshold determinations. 

Ambient noise attenuation with an E-A-RTONE™ Insert Earphone typically exceeds 30 dB in 

the frequency region of 125-8000Hz.  Audiometric air conduction threshold testing down to 0 dB 

Hearing Level (HL) can be reliably performed in the presence of a background noise level not 

exceeding 45 dB(A).  

The current ANSI Standard “Maximum Permissible Ambient Noise Levels“ (MPANLs) For 

Audiometric Test Rooms” (ANSI S3.1-1999 Table 1 and Table 2 ) includes octave band and one-third 

octave band permissible noise levels for both supra-aural and E-A-RTONE™ Insert Earphones.   

The ANSI octave band MPANLs are presented here in Table 2. below for standard earphones and 

for E-A-RTONE™ Insert Earphones with fully inserted E-A-RLINK™ eartips.

ANSI S3.1-1999 - Ears covered: octave band MPANLs using a supra-aural or insert earphone for 

three test frequency ranges re: 20

μ

Pa to nearest 0.5 dB.

2

Eartip connection for 

E-A-RTONE™3A

1

Sound tube 

nipple

Foam eartip

Sound tube

Black 

tubing

Eartip connection for 

E-A-RTONE™5A

E-A-RTONE

Case

Foam eartip

Black 

tubing

Full insertion

Shallow insertion

2-3 mm

3

4

GZXO$

Содержание E-A-RTONE 3A

Страница 1: ...34 8708 1478 6 www 3M EU Safety 3M E A RTONE 3A and 5A Audiometric Insert Earphones GZXO ...

Страница 2: ... output impedance requirements of the audiometric equipment and the E A RTONE Insert Earphone impedance as identified in the Serial number are the same E A RTONE 3A and 5A Insert Earphone replacement cables are not interchangeable between the two models and in the case of the E A RTONE 5A Insert Earphone cable replacement must be both model and impedance specific 5A 10 ohm or 5A 50 ohm E A RLINK f...

Страница 3: ...following procedure is recommended The instrumentation required includes a Sound Level Meter with octave band filter capability and an appropriate condenser microphone calibrated according to the manufacturer s specification A 2cc coupler e g Bruel Kjaer DB 0138 is attached to the microphone The white nylon tube nipple of the E A RTONE 3A Insert Earphone or the stainless steel case nipple of the E...

Страница 4: ... period 3M will repair or replace at its sole discretion any defective part returned to 3M with shipping charges prepaid to the address listed below This shall be the sole and exclusive remedy available for any product defect or failure under this warranty and for any and all claims arising out of the purchase and use of the product and the sole obligation of 3M Company under this warranty shall b...

Страница 5: ...34 8708 1478 6 www 3M EU Safety 3M E A RTONE 3A y 5A Auriculares intraconducto audiométricos E ...

Страница 6: ...intercambiarse entre los dos modelos y en el caso del auricular intraconducto E A RTONE 5A el reemplazo del cable debe ser específico tanto del modelo como de la impedancia 5A 10 ohm o 5A 50 ohm Los filtros auditivos de espuma E A RLINK se diseñan para un solo uso y deben reemplazarse con un juego nuevo por cada prueba realizada La reutilización de estos filtros auditivos podría resultar en una in...

Страница 7: ...imiento siguiente Los instrumentos requeridos incluyen un Sonómetro con capacidad de filtro de banda de octava y un micrófono de condensador apropiado calibrados de acuerdo con la especificación del fabricante Un acoplador 2cc por ejemplo el Bruel Kjaer DB 0138 se conecta al micrófono La boquilla de tubo de nylon blanco del auricular intraconducto E A RTONE 3A o la boquilla de la caja de acero ino...

Страница 8: ... y E A RTONE 5A están libres de defectos en los materiales y mano de obra bajo uso normal durante un período de doce 12 meses desde la fecha de compra al usuario final original aunque en ningún caso se extenderá garantía alguna a más de dieciocho 18 meses a partir de la fecha de fabricación Durante este período de garantía 3M reparará o reemplazará a su sola discreción cualquier parte defectuosa d...

Страница 9: ...34 8708 1478 6 www 3M EU Safety 3M E A RTONE 3A et 5A Écouteurs internes pour audiométrie FzB ...

Страница 10: ...e série sont les mêmes Les câbles de rechange des écouteurs internes E A RTONE 3A et 5A ne sont pas interchangeables entre les deux modèles Pour les écouteurs internes E A RTONE 5A le modèle et l impédance du câble de rechange sont spécifiques 5A 10 ohms ou 5A 50 ohms Les embouts en mousse E A RLINK sont conçus pour une utilisation unique et doivent être remplacés par de nouveaux embouts à chaque ...

Страница 11: ... fabricant Fixez un coupleur 2cc par exemple le DB 0138 de Bruel Kjaer au microphone Le mamelon du tube en nylon blanc de l écouteur interne E A RTONE 3A ou le mamelon du boîtier en acier inoxydable de l écouteur interne E A RTONE 5A est connecté directement au coupleur voir l illustration via une petite portion du canal d aide auditive no 13 Réglez le cadran d intensité de l audiomètre à 70 dB HL...

Страница 12: ... une utilisation normale de douze 12 mois à compter de la date de la vente à l acheteur initial cette durée ne pouvant en aucun cas être allongée à plus de dix huit 18 mois à compter de la date de fabrication Pendant cette période de garantie 3M réparera ou remplacera à sa seule discrétion toute pièce défectueuse renvoyée à 3M avec frais d expédition prépayés à l adresse indiquée ci dessous Il s a...

Страница 13: ...34 8708 1478 6 www 3M EU Safety 3M E A RTONE 3A e 5A Apparecchi endoauricolari audiometrici Iz ...

Страница 14: ...arecchi endoauricolari E A RTONE come indicato dal numero di serie corrispondano I cavi di sostituzione degli apparecchi endoauricolari E A RTONE 3A e 5A non sono intercambiabili tra i due modelli e nel caso degli apparecchi E A RTONE 5A il cavo di sostituzione deve indicare esattamente sia il modello specifico che la sua impedenza 5A 10 ohm o 5A 50 ohm I tappi in schiuma E A RLINK sono monouso e ...

Страница 15: ...on capacità di filtro a banda di ottava e un appropriato microfono a condensatore calibrato in base alle specifiche del produttore Al microfono deve essere collegato un accoppiatore 2cc ad esempio Bruel Kjaer DB 0138 Il raccordo del tubo di nylon bianco degli apparecchi endoauricolari EARTONE 3A o quello in acciaio inossidabile degli apparecchi endoauricolari EARTONE 5 deve essere collegato dirett...

Страница 16: ...all acquirente originale in nessun caso la garanzia dovrà eccedere i diciotto 18 mesi dalla data di fabbricazione Durante questo periodo di garanzia 3M si impegna a riparare o sostituire a propria discrezione qualsiasi parte difettosa restituita a 3M con spese di spedizione prepagate all indirizzo specificato di seguito Ciò costituirà il solo ed unico rimedio offerto per qualsiasi difetto o guasto...

Страница 17: ...34 8708 1478 6 www 3M EU Safety 3M E A RTONE 3A und 5A Audiometrischen Stöpsel Kopfhörer Dza ...

Страница 18: ... durch die Seriennummer angegeben identisch sind Die Ersatzkabel für die E A RTONE 3A bzw 5A Stöpsel Kopfhörer sind zwischen beiden Modellen nicht austauschbar Beim Kopfhörer Modell E A RTONE 5A muss das Ersatzkabel nicht nur mit dem Modell übereinstimmen sondern auch die spezifische Impedanz aufweisen 5A 10 Ohm oder 5A 50 Ohm E A RLINK Schaumstoff Stöpsel sind für den Einweg Gebrauch bestimmt und...

Страница 19: ...rliche instrumentelle Ausstattung umfasst einen Geräuschpegelmesser mit Oktavband Filterkapazitat sowie ein entsprechendes gemäß Spezifikation kalibriertes Kondensierungsmikrophon Ein cc Koppler z B Bruel Kjaer DB 0138 wird an das Mikrophon angebracht Der weiße Nippel des Nylonschlauchs der E A RTONE 3A Stöpsel Kopfhörer oder der Edelstahlnippel des Gehäuses der E A RTONE 5A Stöpsel Kopfhörer wird...

Страница 20: ... von Materialdefekten oder Herstellungsfehlern sind In keinem Fall wird eine Garantie von mehr als achtzehn 18 Monaten ab dem Herstellungsdatum übernommen Während dieser Gewährleistungsfrist kann 3M in freiem Ermessen alle zurückgegebenen schadhaften Teile reparieren oder aber ersetzen Die Versandkosten sind im Voraus an die unten angegebene Adresse zu entrichten Dies ist die einzige und ausschlie...

Отзывы: