49
10.2
System
(system
parameter) setting item
Table 10 - 2System (system parameters) setting items (default password 0200)
Set item
Description
Options /
value range
Default
value
explain
New Password
New
password
0
~
9999
0200
Signal
Signal input
Qmax
Scale flow
(m3/h)
Change
according to
caliber setting
Change
according to
caliber
setting
Range
setting
affects
flow
percentage, current output, etc. The
unit is fixed as m3/h. The decimal
point can be set by pressing ↓.
Low Cutoff%
Small signal
removal (%)
0
~
9.9
1
It is used to calculate a threshold.
When the traffic is less than this
threshold, the traffic will be cut off. It
displayed first, and the decimal
digits will be reduced accordingly.
Total Units
Cumulative
flow unit
m3,L,mL
M3
The cumulative flow data is
displayed
after
conversion
according to this configuration.
Totalizer
decimals
Cumulative
decimal point
1,2,3
3
The decimal digits of the cumulative
flow data value are displayed
according to this configuration, but
if the integer digit is too long, the
integer digit will be displayed first,
and the decimal digits will be
reduced accordingly.
Power
Units
Power unit
kW,MW,kJ/h,MJ/h
,GJ/h,cal/h,kcal/h,
Mcal/h
KW
The power data is displayed after
conversion according to this
configuration.
Power
decimals
Power decimal
point
1,2,3
3
The decimal digits of the power
data value are displayed according
to this configuration, but if the
integer digit is too long, the integer
digit will be displayed first, and the
decimal digits will be reduced
accordingly.
Energy
decimals
Energy
decimal point
1,2,3
3
The decimal digits of the heat data
value are displayed according to
this configuration, but if the integer
digit is too long, the integer digit will
be displayed first, and the decimal
digits will be reduced accordingly.
Damping(s)
Flow damping
time (s)
0
~
99
3
The flow display value is affected by
the damping value. The greater the
damping, the slower and more
stable the response is; The smaller
the damping, the faster the
response, but the more unstable.
T
Damping(s)
Temperature
damping time
(s)
0
~
99
4
The temperature display value is
affected by the damping value. The
greater the damping, the slower
and more stable the response
is; The smaller the damping, the
faster the response, but the more
unstable
User No.
User number
0
~
99999999
0