Recommended Minimum Preventive Maintenance Schedule
9650-000820-01 Rev. K
Propaq M Operator’s Guide
17-5
Recommended Minimum Preventive Maintenance
Schedule
Operational Tests should be performed at regular intervals. The Operational Tests augment the
automated self-tests that the Propaq M unit performs to help ensure readiness. The following
Operational Tests may be performed by the user:
•
Keypad Test
•
LED Test
•
LCD Test
•
RFU Indicator Test
•
Audio Test
Additional Tests may be performed by a qualified Biomedical Technician (BMET):
•
Battery Test
•
Fan Test
•
NIBP Test
•
USB Test
The Operational Test Procedures are described in the
Propaq M Service Manual
.
In addition to inspection and cleaning instructions, the following is recommended:
•
Perform NIBP calibration check
•
Perform CO
2
calibration check
•
Perform Functional Verification Tests
These tests shall be performed every year or according to local requirements (to be performed
by a qualified Biomedical Equipment Technician (BMET)) as described in the
Propaq M
Service Manual
.