Temperature System Messages
9650-000820-01 Rev. K
Propaq M Operator’s Guide
12-5
You can select one of the following labels for each temperature channel:
If you don’t select a label, the Temperature channels appear with the default labels of T1 and
T2.
Temperature System Messages
The Propaq M unit may display the following messages when monitoring Temperature.
Note:
The temperature function performs a self test when initially powered on and also
performs system tests automatically, every 10 seconds, while this function is active.
Source Label
Description
ART
Arterial temperature probe
CORE
Central body temperature or tympanic membrane temperature
probe
ESOPH
Esophageal temperature probe
RECT
Rectal temperature probe
SKIN
Skin temperature probe (surface application)
VEN
Ventilator airway temperature probe
NASO
Nasopharynx or nasal/oral temperature probe
System Message
Cause
CHECK PROBE
The temperature probe is disconnected. Check probe
and reconnect it.
PROBE FAILURE
The temperature probe is defective. Replace the
Temperature probe.
TEMP DISABLED
A system error has occurred. The Propaq M unit cannot
take temperature measurements and should be serviced.