Enabling/Disabling SpO2 Alarms and Setting Alarm Limits
9650-000820-01 Rev. K
Propaq M Operator’s Guide
10-13
Setting Upper and Lower SpOC Alarm Limits
The following table lists the default SpOC limits for adult, pediatric, and neonate patients, and
gives the range in which you can set these limits
Setting Upper and Lower PVI Alarm Limits
The following table lists the default PVI limits for adult, pediatric, and neonate patients, and
gives the range in which you can set these limits
Setting Upper and Lower PI Alarm Limits
The following table lists the default PI limits for adult, pediatric, and neonate patients, and
gives the range in which you can set these limits
Patient Type
SpOC Limit Default
SpOC Limit Range
Adult
Lower: 10 ml/dL
Upper: 25 ml/dL
Lower: 0 - 34.9 ml/dL
Upper: 0.1 - 35 ml/dL
Pediatric
Lower: 10 ml/dL
Upper: 25 ml/dL
Lower: 0 - 34.9 ml/dL
Upper: 0.1 - 35 ml/dL
Neonate
Lower: 10 ml/dL
Upper: 25 ml/dL
Lower: 0 - 34.9 ml/dL
Upper: 0.1 - 35 ml/dL
Patient Type
PVI Limit Default
PVI Limit Range
Adult
Lower: 5%
Upper: 40%
Lower: 0
-
98%
Upper: 2 - 100%
Pediatric
Lower: 5%
Upper: 40%
Lower: 0
-
98%
Upper: 2 - 100%
Neonate
Lower: 5%
Upper: 40%
Lower: 0
-
98%
Upper: 2 - 100%
Patient Type
PI Limit Default
PI Limit Range
Adult
Lower: 0%
Upper: 20%
Lower: 0
-
19.8%
Upper: 0.2 - 20%
Pediatric
Lower: 0%
Upper: 20%
Lower: 0
-
98%
Upper: 2 - 100%
Neonate
Lower: 0%
Upper: 20%
Lower: 0
-
98%
Upper: 2 - 100%