Cleaning / Disinfection
10
Page 13
Before starting any maintenance and cleaning measures, the device must always be
turned off with the main switch, and the mains cable must be disconnected.
Make sure that during the cleaning and disinfection the labelling of the device (such
as warnings, label of control devices and identification plate) are not damaged.
Make sure that no liquid enters the instrument, foot switch or applicator during
cleaning or disinfection. Do not use sprays.
If liquid enters into the device or applicator during the cleaning or disinfection
process, please turn off the device, prevent it from getting used again, and contact
your service representative.
The device and its application device are considered as uncritical in relation to
hygiene due to the use on uninjured and healthy skin.
Housing / footswitch
Cleaning: In the event of visible contamination, the housing, footswitch and all
cables can be cleaned using commercially available soft alcohol-free plastic
cleaners. Wipe the surface until the dirt is removed, using a soft cloth soaked
according to the specifications of the manufacturer of the cleaning agent but not
dripping.
Disinfection: Housing, footswitch and all cables can be disinfected using disinfectant
wipes. To do so, use a commercially available alcohol-free disinfectant for metal and
plastic, with bactericidal, viricidal and fungicidal properties. Observe the application
instructions of the manufacturer. Wipe all surfaces using a cloth soaked according to
the specifications of the manufacturer of the disinfectant, but not dripping, or use pre-
impregnated disinfectant wipes (so-called wipes).
If applicable, also observe requirements for drying or post-cleaning.
Applicator head /
handpiece
Cleaning: Remove the silicone protection cap from the applicator before cleaning.
Then proceed as indicated under “housing/footswitch”.
Disinfection: Remove the silicone protection cap from the applicator head before
disinfecting. Then proceed as indicated under “housing/foot switch”.
Caution:
If flammable solutions are used for cleaning and disinfection, solutions
must be allowed to evaporate before using the device. Otherwise, an inflammation
may arise!
Note
:
Use the device only in hygienic environment.
Caution
!
When using lubricants, the silicone protection cap must be placed over the applicator
to avoid contamination.
If you are working without a protection cap, the lubricant may enter into the interior of
the applicator and the handpiece, which can lead to permanent contamination and
malfunction of the
handpiece.
Note
:
In this case, the warranty will expire.
Summary of Contents for Z Wave Q
Page 1: ...User Manual ZWaveQ USA...
Page 2: ......
Page 3: ...Illustrations Front of the device Fig 1 Encoder Handpiece...