Manufacturer´s EMC declaration
16
Page 27
Table 2
Guidance and Manufacturing Declaration- Electromagnetic Immunity
The device ZWave
Q
is intended for use in the electromagnetic environment specified below. The customer or user of
the device ZWave
Q
should ensure that it is used in such environment.
Immunity Tests
IEC 60601- Test Level
Compliance Level
Electromagentic Environment -
Guidlines
Elctrostatic
Discharge (ESD)
in accordance
with IEC 61000-
4-2
± 8 kV Contact Discharge
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV Air Discharge
± 8 kV Contact Discharge
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV Air Discharge
Floors should be made from wood,
concrete or ceramic tiles. If floor are
covered with synthetic material, the
relative humidity must be at least 30
%
Electrical fast
transient/ burst
in accordance
with IEC 61000-4-
4
± 2 kV
100 kHz repetition
frequency
± 2 kV
100 kHz repetition frequency
The supply voltage quality must
correspond to that of a typical
commercial or hospital
environment.
Surges in
accordance with
IEC 6100-4-5
-Line-to-Line-
± 0,5 kV, ± 1 kV
± 0,5 kV, ± 1 kV
Surges in
accordance with
IEC 6100-4-5
-Line-to-Earth-
± 0,5 kV, ± 1 kV, ± 2 kV
± 0,5 kV, ± 1 kV, ± 2 kV
Voltage dips in
accordance with
IEC 61000-4-11
0 % U
T;
0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % U
T;
0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
The supply voltage quality must
correspond to that of a typical
commercial or hospital environment.
If the user of the device ZWave
Q
requires continued operation, even in
the case of interruptions in the power
supply, it is recommended that the
device ZWave
Q
be powered from an
uninterrupted power supply or a
battery.
0 % U
T;
1 cycle and 70%
U
T;
25/30 cycles
Single phase: at 0°
0 % U
T;
1 cycle and 70% U
T;
25/30 cycles
Single phase: at 0°
Voltage
interruptions
accordance with
IEC 61000-4-11
0% U
T;
250/300 cycle
0% U
T;
250/300 cycle
Magnetic filed of
supply frequency
(50/60 Hz) in
accordance
with IEC 61000-
4-8
30 A/m
50 Hz oder 60 Hz
30 A/m
50 Hz
Magnetic fields at mains frequency
should have the typical values found
in a business or hospital
environment.
Note: U
T
is the mains AC Voltage before application of the test level