Technical Information
13.4 VacoS
13
Page 63
Mains voltage
100 – 120 V / 50 Hz / 60 Hz (1)
230 – 240 V / 50 Hz / 60 Hz
220 V / 60 Hz (2)
Mains fuse
2 x T 0.4 AL, 100 – 120 V, 5 x 20 mm
2 x T 0.2 AL, 230 – 240 V, 5 x 20 mm
Electrical energy
consumption
max. 50 VA
Protection class
I
Appli
ed
P
art
Type BF
Operating mode
Intermittent operation: 30 minutes on, 10 minutes off
Dimensions
322mm x 234mm x 130mm
Weight
3.0 kg
Operation
+10°C to +35°C, 20% to 80% relative humidity without condensation,
700 hPa – 1060 hPa
Storage and transport -10 °C to +50 °C, 10% to 90% relative humidity without condensation,
at 700 hPa - 1060 hPa
Negative pressure
12-60 kPa
Pulse operation
Period adjustable to 1 to 8 seconds in steps of 0.5 seconds, ratio 1: 1
Accuracy
± 20%
Appli
ed
P
art
Vacuum suction cups and sponges
Note:
Storage and transport in original packaging only.
Subject to technical modifications!