7 Device Set up
7.2 Installation of cables for VacoS
7
Page 15
Note:
The control and operation of VacoS is via Soleo SonoStim/Soleo Galva. A stand-
alone operation is not possible.
Preparation
Position Soleo SonoStim/Soleo Galva so that the devices are aligned with each
other. The sockets of the Vaco electrodes hoses must be on the front (fig. 4).
Connection
of VacoS with
SonoStim/Galva
Make sure when mounting the connecting cable that when connecting the cable to
Soleo SonoStim/Soleo Galva (16+18), the green arrow is in the plug below.
Make sure when mounting the connecting cable that when connecting the cable to
VacoS (28+30), the green arrow is in the plug below.
Channel I
Connect the connecting cable to the intended port on Soleo SonoStim / Soleo
Galva (18) and on VacoS (30).
Channel II
Connect the connecting cable to the intended port on Soleo SonoStim / Soleo
Galva (16) and on VacoS (28).
Note:
If Soleo SonoStim / Soleo Galva is operated in conjunction with VacoS, the
connection is made with the electrode cable on VacoS. The assembly is described
below.
Assembly
Electrode cable
Make sure when mounting the electrode cable for VacoS that the green arrow is
facing upward in the insertion direction.
•
Connect the electrode cable for channel I to the appropriate socket (29) on
VacoS.
•
Connect the electrode cable for channel II to the appropriate socket (27) on
VacoS.
•
Insert the red alligator clips to the red connector of the electrode cable.
•
Insert the black alligator clips to the black connector of the electrode cable.