Soleo SonoStim – Instructions for
operation
9.1 Electrotherapy
9
Page 28
(6) Status line
The status bar shows information regarding the current therapy status. If therapy is
not active, the "Ready" display appears; when therapy is in progress the text "Active"
appears.
(7) Title bar
In the title bar, the name of the current ultrasound therapy program is displayed.
(8) Effect
Gives an overview of the medical effects of the current flow form.
Note:
If the parameters of the waveform (pulse time or pause) change, the current form's
effects will change. Therefore, after a change the window is no longer visible.
(9) Threshold
Represents the threshold parameters of the current treatment programme in the
activated state. The parameters of increase time, hold time and break time are
displayed.
If there is no threshold activated, the threshold window is shown minimised.
Activation of the "threshold" button opens the "threshold parameters" window. Here
threshold can be activated or the parameters can be user-defined.
Changed parameters are taken over by activating the "OK" button and the "Cancel"
button aborts the process.
Note:
If no threshold is provided for a programme, the window is completely hidden.
(10) Parameters
Represents the parameters of the current therapy programme.
These are the pulse and pulse pause time (and hence frequency), polarity, CC / CV
and galvanic share parameters.
Activation of the "parameters" button opens the "current form parameters" window.
The parameters can be user-defined.
Use the arrow keys to change pulse parameters or the values within the current form
up and down.
Change the polarity, change CC/CV and galvanic share by activating the
corresponding button.
Note:
Modes 1.3 to 1.5 are not useful for all types of current and are therefore not always
displayed.