Function test
12
41
en
Puls
runs a self-test that checks all internal components after it is
switched on.
An error message is shown in case of faults.
In addition, a function test shall be made as follows.
This test shall be made monthly or in case of doubt about the proper
function of the device.
Note:
Before performing the function test, check whether the handpiece and
the footswitch are connected correctly to the device.
Check for proper mains connection.
Function test
Testing
Switch on the device.
Depress the footswitch briefly – the fan and shockwave generator will
start immediately, whereby the shockwave generator has to operate at
the frequency indicated on the display (5 Hz as default value).
Note:
On conclusion of the test, switch off the device at the main switch.
If a treatment is to be performed immediately afterwards, set the required
treatment parameters and proceed as mentioned in Chapter 4.
Summary of Contents for enPuls
Page 1: ...Operating Instructions enPuls Version 2 0 GB...
Page 2: ......