Maintenance
10
38
Monitoring the
handpiece
temperature
Generating mechanical shockwave energy causes a considerable build
up of heat in the handpiece. To avoid shortening the lifetime of the
handpiece, a temperature switch has been integrated. This triggers an
internal switch-off, if the temperature becomes too high, forcing the
handpiece to cool.
If the temperature switch is activated, this is indicated by a message on
the display and shocks can no longer be emitted.
After acknowledging the message with ‘OK’, the treatment screen comes
to the foreground with the message 'Over temperature' in the status bar.
As soon as the handpiece has reached the operating temperature, the
message 'Over temperature' is replaced by the message 'Ready' in the
status bar and the treatment can be continued.