Error Messages
13
42
In the status bar the
message 'Handpiece not
found' appears.
Check that the handpiece is correctly connected
Monitoring
the handpiece
temperature
Generating mechanical shockwave energy causes a considerable build
up of heat in the handpiece. To avoid shortening the lifetime of the
handpiece, a temperature switch has been integrated. This triggers an
internal switch-off, if the temperature becomes too high, forcing the
handpiece to cool.
If the temperature switch is activated, this is indicated by a message on
the display and shocks can no longer be emitted.
After acknowledging the message with ‘OK’, the treatment screen comes
to the foreground with the message 'Over temperature' in the status bar.
As soon as the handpiece has reached the operating temperature, the
message 'Over temperature' is replaced by the message 'Ready' in the
status bar and the treatment can be continued.
In the status bar the
message 'Ready' appears
and despite depressing
the footswitch no shock
is triggered.
Check that the shock energy is set.
Check that the footswitch is correctly connected.
Inspect the footswitch cable for any damage or kinks.
Check whether the footswitch dome can be moved or whether it is
blocked.
Please contact after-sales service if this fails.
After-sales service is reached through your authorised sales
representative or by contacting the head office in Neu-Ulm.
No SD-Card found
If the SD card is not inserted, the message ‘SD card not found’ appears
when the 'Favourites' and 'Memory' buttons are pressed.
Insert card and confirm with ‘OK’.
Disposal
The device must be disposed of by an approved disposal company and
must not be discarded with household or special waste.