Troubleshooting
11
39
Failure or malfunction
of the handpiece
Check to ensure that the handpiece plug is properly connected to the
device.
It must be fully engaged.
Check the cable of the handpiece for any mechanical damage.
Irregular delivery
of shockwaves /
overheating
of handpiece
Possible cause 1:
Wear of applicator head
Difficult to move due to wear
Applicator heads are wear parts and should be replaced after a specific
number of shocks.
Remedying
cause 1
Removal of parts subject to abrasion:
Remove the applicator head from the handpiece and clean the rear
dome thoroughly. Then hold the handpiece, without the applicator head,
with the opening downward and, at 2 or 5 Hz frequency, release a few
shocks (maximum 10) at the lowest energy level. Then reinsert the
applicator head.
If the error still occurs, the applicator head has to be changed.
Possible cause 2: Wear of shockwave generator
The shockwave generator is an expendable part and should be replaced
after 2 million shocks.
Check the total number of shocks of the device in the configuration
menu.
Remedying
cause 2
If the total number of 2 million shocks has been reached or exceeded,
the shockwave generator must be replaced.
To replace the shockwave generator, contact a qualified customer
engineer or the head office in Neu-Ulm, Germany.