Troubleshooting
11
40
No response at
main switch /
display remains
dark
Make sure that the mains plug is properly inserted in the power outlet
and the device connector is firmly plugged into the device port.
Inspect the mains cable for damage.
Check the power supply and the power plug.
Above the mains input socket of the device, there are fine-wire fuses,
which isolate the mains voltage in the event of any electrical problem.
Open the flap and check the fuses.
Replace any faulty fuses.
Only replace a fuse with one of exactly the same name or one that is
equivalent. Before doing this, check the entire power supply for any
possible faults.
If the error occurs again, it is essential to inform the service/after-sales
service department.