The Zeroplus Logic Analyzer
Installation Guide
The Zeroplus Logic Analyzer Installation Guide
Page 22
Fig 3-4 – Trigger Properties
Step 2.
Trigger Level Setup
Click the drop down menu of Trigger Level on Ports A , B , C and D to select
the Trigger Level as the voltage level that a trigger source signal must reach
before the trigger circuit initiates a sweep.
Tip:
There are four commonly used preset voltages for Trigger Level, TTL,
CMOS (5V), CMOS (3.3V), and ECL. Users also may define their own
voltage from -6V to +6V to fit with their DUT.
Port A represents the pins from A0 ~ A7 on the signal connector of the Logic
Analyzer, and so do Ports B, C and D. The voltage of each port may be
configured independently.
Fig 3-5 –
Trigger Properties Error
Step 3.
Trigger Count.
Type the numbers or select the number from the pull down menu of the Count
on the Tool Bar or Click the pull down menu of the Trigger
Count on the Trigger Properties dialogue as shown in Fig 3-5
The system will be triggered where the Trigger Count is set as shown in
Figs 3-6,3-7 and Fig 3-8.
Summary of Contents for LAP-321000U-A