System Specifications
This product may contain remanufactured parts or parts that have had incidental use, all of which are
equivalent in performance to new parts.
Essential Performance
Table 22: Performance Characteristics
Device State
When Target Is:
Required Device Performance Is:
Cooling
Below 5°C
No more than 1°C below the target temperature
Heating
Above 30°C
No more than 1°C above the target temperature
Steady state
2-10 InHg
Vacuum pressure controlled to within ± 0.5 inches of Hg
Environmental Requirements
The system and its components are designed to operate normally when stored, shipped, and operated
under the conditions specified in Table 23: Environmental Requirements.
Use of the system in an oxygen-rich environment may cause fire. Do not use the system in an
oxygen-rich environment.
The system may not operate as expected if it is stored or operated in conditions of excessive
heat, humidity, or atmospheric pressure. Operate and store the system in a room that meets
the stated requirements.
Table 23: Environmental Requirements
Condition
Shipping/Storage Requirement
Operating Requirement
Temperature
14°F to 140°F (-10°C to 60°C)
59°F to 82°F (15°C to 28°C)
Humidity
10% to 95% (non-condensing)
10% to 70% (non-condensing)
Atmospheric
pressure
14.7 psi (101.33 kPa) to 10.1 psi (69.64
kPa).
14.7 psi (101.33 kPa) to 10.1 psi
(69.64 kPa).