Care and Maintenance
S100 / OPMI pico
Version 7.0
Page 228
G-30-1781-en
Changing the fuse - halogen illumination system
The fuses are integrated in the power inlet socket for the power cable. To
change a fuse, proceed as follows:
• Turn off the unit at the power switch.
• Unplug the power cable from the line power socket.
• Unplug the power cable from the power inlet socket.
• Insert a suitable object under two flaps (1) and pull out fuse holder (2).
• Remove the defective fuse.
• Insert the new fuse and push the fuse holder back into its compartment.
Note the voltage rating of the line fuse. This rating can be found on the type
label.
• Connect the device to the power supply.
• Turn on the device at its power switch.
Fig. 103: Changing the fuses
CAUTION
Use suitable device fuse
• Only use the fuse indicated for the device (see order data).
Fuse rating
For 100 V...125 V~ T 6,3 A/H 250 V
For 220 V...240 V~ T 3,15 A/H 250 V
1
2
1
Summary of Contents for S100/OPMI pico
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...