Preparations for use
S100 / OPMI pico
Version 7.0
Page 108
G-30-1781-en
Connections
Connecting the S100 floor stand
• Ensure that the rated voltage indicated in the display window (3) (only for
halogen/xenon) corresponds with the rated voltage for the network at the
site of the installation.
• Ensure that the unit has been switched off at the power switch before
plugging in or unplugging connectors at the power input socket (4).
• Make sure that power switch (1) is OFF.
• Plug the supplied power cord on the supply unit into the power inlet sock-
et (4).
• Connect the power cord via a socket to the power supply system.
CAUTION
Risk of electric shock caused by inadmissible or defective accessories!
Inadmissible or defective accessories may lead to increased leakage current!
• Never touch any power outlets or video ports during patient contact.
• When setting up your system, make sure to comply with the require-
ments of IEC 60601-1:2005, chapter 16; in some countries IEC 60601-1-
1:2000 may still be valid.
• Also comply with the Medical Device Operator Ordinance or other na-
tional codes.
NOTE
Risk of tripping!
Inappropriate routing of cables increases the risk of tripping.
• Always run cables in a manner that will not impede the user's move-
ments.
CAUTION
Risk of electric shock
Risk of electric shock for operator and patient if cables and plug connections
loosen and if housing parts are under voltage.
• Only connect the power cord using a socket with proper protective earth
connection.
Summary of Contents for S100/OPMI pico
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...