S100 / OPMI pico
Preparations for use
Mounting the MORA interface
• Give securing screw (3) a few turns to loosen it.
• Remove the dust cover and store it in a safe place.
• Place MORA interface (7) on the surgical microscope and firmly tighten se-
curing screw (3).
• Insert binocular tube (5) into the mount of MORA interface (7) and tighten
securing screw (6).
• Tilt the microscope body by approx. 25° to the left or right and lock it in
the required position using friction screw (2).
• Set ring diaphragm (4) to meet your specific requirements.
Fig. 52: Mounting the MORA
interface
5
1
approx. ± 25°
3
2
4
6
7
3
Summary of Contents for S100/OPMI pico
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...