1 About these Installation Requirements | 1.1 Further Applicable Documents

ZEISS

1 About these Installation Requirements

This document describes the location requirements for the installation of the LSM 900. Here-
inafter referred to as Microscope System.

It helps to determine a suitable location within facilities and to prepare the location before deliv-
ery of the Microscope System. The suitability of a location depends on various factors such as:

§

Room size

§

Floor stability

§

Environmental conditions

§

Availability of supplies

The room size should be able to accommodate the Microscope System as well as potential acces-
sories and 3rd party systems that may be attached to the Microscope System. For servicing the
equipment, adequate access space is also required. The selected location should conform with the
correct environmental conditions to ensure the safe and ideal operation of the Microscope System
and its application whilst still ensuring the safety and comfort of the operators. Environmental
conditions are often influenced by surrounding installed systems or activities. We can distinguish
between:

§

Vibrations

§

Acoustic noise

§

Room temperature

§

Air quality

The customer is responsible for ensuring that the conditions specified in this document are met
and that the required equipment is available before delivery and installation. A ZEISS service repre-
sentative will gladly assist you in examining the installation conditions and suitable measures.

1.1 Further Applicable Documents

Please take also note of the following documents:

Downloads

Brochures

, Safety data sheets, ISO certificates, CSA certificates and EU declarations of conformity

as well as software and firmware updates can be found at 

https://www.zeiss.com/microscopy/int/

downloads.html

 or ask your ZEISS Sales & Service Partner.

Local and National

health and safety

regulations

Local and national health and safety regulations should be adhered to for the site and the use of
the system. Please consult with your ZEISS Sales & Service Partner in the event of these regulations
conflicting with the installation requirements of the Microscope System.

System and 3rd

Party Components,

Accessories

The Microscope System can be configured in many ways. Information about the individual com-
ponents, enhancements and accessories can be obtained from your ZEISS Sales & Service Partner.
Also refer to the 3rd Party documentation of the manufacturer.

4

Installation Requirements ZEISS LSM 900 | en-US | Rev. 2 | 349500-8183-000

Summary of Contents for LSM 900

Page 1: ...Installation Requirements ZEISS LSM 900 Laser Scanning Microscope...

Page 2: ...archiving purposes shall remain unaffected thereby Any viola tions may be prosecuted as copyright infringements The use of general descriptive names registered names trademarks etc in this document d...

Page 3: ...rvoltage Category 10 2 3 2 UPS Requirements 12 2 4 Network Connectivity 12 2 4 1 ZEISS Predictive Service 13 2 5 Safety and Radiation Information 14 2 5 1 Laser Safety 14 2 5 2 Biosafety Level of the...

Page 4: ...oom temperature Air quality The customer is responsible for ensuring that the conditions specified in this document are met and that the required equipment is available before delivery and installatio...

Page 5: ...than 50 dB Z class 200 to 500 Hz less than 55 dB Z class 2 1 2 Air Conditioning and Quality The following nominal power usage for the Microscope System and possible accessories is shown in Watts To m...

Page 6: ...porary conductivity caused by condensation is to be expected Example A circuit used in an of fice environment The circuitry inside a computer would fall in this cate gory 3 Conductive pollution occurs...

Page 7: ...d in closed rooms It is recommended to install the Microscope System in a dimmed room where artificial illumination sunlight or other light sources cannot interfere with image acquisition The Microsco...

Page 8: ...depending on the options purchased with the system For possible changes in preferential layout please dis cuss with your ZEISS representative The placing of the user table can vary depending on avail...

Page 9: ...ds and supplied socket with protective ground contact securely connected The provided electrical connection must be in accordance with the applicable electrical codes for the country of installation I...

Page 10: ...eferred to as Installation Category to address transient voltages Category IV devices can handle the largest transients relative to the normal working voltage Category I devices can handle only small...

Page 11: ...oltage levels of these loads must be rated at this overvoltage category III In fixed installations and for cases where the reliability and the availability of the equipment is subject to special requi...

Page 12: ...til normal mains is restored Please ensure to switch off the Microscope System be fore the batteries run low in a blackout In areas where the mains supply is very unstable or not very clean the most s...

Page 13: ...re the above network connectivity and predictive service will oper ate under the parameters shown in the table below Ports 443 Protocols HTTPS secure websocket URL of ZEISS server predictive service z...

Page 14: ...s the responsible body the operator is responsible for complying with laser safety regula tions The following types of lasers are currently intended for use Using other lasers than indicated be low re...

Page 15: ...ever some do exist in a small number of places in the US and around the world As the highest level of biological safety a BSL 4 lab consists of work with highly dangerous and exotic mi crobes Infectio...

Page 16: ...e user if the air supplied is from the building facilities or from gas bottles Due to acoustic noise and vibrations where possible the compressor should be installed in a sep arate room close to the w...

Page 17: ...f the system where necessary Some components such as the tables are large heavy or bulky and may require extra assis tance to get the units into the allocated site Maximum shock resistance Do not drop...

Page 18: ...s used not possi ble for Airyscan Anti vibration plate for Axio Imager 320 300 45 7 Anti vibration plate for Axio Observer 800 525 45 7 Microscope Axio Examiner Z1 700 390 820 24 Axio Imager Z2 560 39...

Page 19: ...EU RoHS and 2012 19 EU WEEE as applicable for the product In addition to the European and international guidelines and standards the 21 CFR 1040 10 Performance Standards for light emitting products la...

Page 20: ...ch requirements are yet to be completed ZEISS can be requested to assist with the checking of the environmental conditions Environmental conditions in and around a system can change with time due to v...

Page 21: ...adiation Information X Ray Laser Safety Biosafety Level of the Laboratory Cooling Water Supply Cooling liquid and or chemicals Exhaust Line Gas Supplies Compressed Air Nitrogen Additional Gas Transpor...

Page 22: ...l Zeiss Promenade 10 07745 Jena Germany phone 49 1803 33 63 34 fax 49 3641 64 3439 info microscopy de zeiss com www zeiss com microscopy Installation Requirements ZEISS LSM 900 en US Rev 2 Modificatio...

Reviews: