background image

40x and REAR ONE 100x; with objective 12,5x is only a large ring watch) the small bright ring is visible: strict phase contrast 

(with objectives 25x, 40x and REAR ONE 100x) if the picture of the aperture diaphragm centrically to Bildder ring screens is 
not appropriate, for aperture diaphragm with centering screws place behind (fig. 1/9).

  

Bertrandlinse switch off. In the field appears the picture of the object in the phase contrast, which is improved by connecting the 

green of filter.

  

The instruction in a general manner following the appropriate ring screens are centered to the remaining, in the context of the 
microscopic intention intended objectives. The individual ring screens are independently shifted. After unique centring objectives and 
ring screen guns can be usually switched without further centring expenditure.

  

Quick change between phase contrast and quasi - light field is possible:

  

by full opening of the aperture diaphragm (with objectives 25x, 40x and REAR ONE 100x),

 

through scolded the ring screen gun on symbol

 

.

  

Operation of the small mechanisms for phase contrast

  

Objectively and/or phv into the lens turret screw pH in and scolded into working position. Is to be worked alternatively with a 

small mechanism DIK D, an empty intermediate ring must between phase objective and lens turret (with white marking; 
Intermediate ring is a component of the small DIK mechanism) to be screwed in, in order the off lengthens the objectives to 
secure.

  

Ring screen into the single screen admission insert, objektivwärts pointing inscription of the ring screen; Finger forces away 

avoids.

  

Preparation present, focus and after Köhler light up favourably focus (with difficulty visible objects with to a large extent 

geschlosse ner aperture diaphragm. Afterwards again fully open aperture diaphragm).

  

Bertrandlinse switch on focus (

 (5)) and on (grey brighten up bright underground appearing) the phase rings of the objective.

  

Single screen admission into the appropriate swallow guidance of the Kondensorsinhängers to attack push in and with knurling 

tool mother (fig. 3/34) wedge. In the objective pupil the picture (grey) of the phase rings is overlaid now by the picture (black) 
of the Ringblen de.

  

Aperture diaphragm so far close, to their edge becomes visible and with centering screws (fig. 1/9) to the phase rings to center.

  

Bring the picture of the lighting ring with the help of the centering screws at the single screen admission with the picture of the 

phase rings to the cover.

  

Following work procedures in a general manner as during mechanism for phase contrast GF-PA.

  

(12) Differential interference contrast (DIK d) with the standard equipment without Kontrasttubus _

  

Information about the procedure and about the delivery offer of DIK mechanisms, recommendable equipment, advantages and possibilities 
of the small mechanisms DIK D find you in block letters No. 30-G0540

  

Operation of the mechanism for differential interference contrast GF-PA _

  

Intermediate rings of the mechanism DIK D onto the objectives GF-PA of the standard equipment screw; the allocation 

intermediate ring objective goes out of the imprinted en größerung out (the intermediate rings DIK D 12.5/25 to function both 
with GF-PA 12,5x as well as with GF-PA 25x).

  

The combinations intermediate ring objective into the objective RH screw in. The objective 12,5x in at the gun with 1 marked eye, 

remaining objectives in the number of rising enlargements into the eyes 2, 3, 4… screw in a general manner favourably. This 
allocation even with later work with the DIK mechanism maintained, in order adjusting spends to avoid.

  

Prism guns into the appropriate swallow guidance of the condensor lens A TRAILEr to attack push and with knurling tool mother 

in (fig. 3/34) wedge. Prism guns scolded in such a way that in the sight the symbol appears

.

  

Summary of Contents for JENAVAL

Page 1: ...of this block letters can occur The rendition also in part is permitted only with our permission The right of the translation we reserve ourselves For publications we make reproductions available of t...

Page 2: ...he gun mark that is to be read off which modulator in the path of rays is Adjust Linkage for photo exit fig 4 37 push in Shift levers for lens 1 1 6 fig 4 41 on symbol 1x flat ten Contrast loop switch...

Page 3: ...so freely selected filters The photographs in the filter change over switch 5x are furnished to one for filter thick up to 4 mm in the filter change over switch 10x are intended the gun for the admiss...

Page 4: ...e g Thick ones of the slide l to 0 mm and on microscope table for observation furnish Set screw fig 3 31 on approximately 10 mm of length out screw Annunciator screen picture focus and center Set scr...

Page 5: ...2 5x _____________ For illuminating these fields the overview condensor lens aplanatic lens 0 12 is switched on fig 3 32 Preparation with middle drying system focus Annunciator screen picture with con...

Page 6: ...ded is if with the objective GF PA 1x 0 03 spec were to be worked This objective results in gun balanced system for a Objektfeld of 25 mm in diameter together with the lens 1 1 6 Focus the picture wit...

Page 7: ...V input voltage indicates the scale line to 6 exactly 6 V In the range between 4 up to 6 V is to be counted on deviations up to 0 3 V 11 Phase contrast with the standard equipment without Kontrasttub...

Page 8: ...Bertrandlinse switch on focus 5 and on grey brighten up bright underground appearing the phase rings of the objective Single screen admission into the appropriate swallow guidance of the Kondensorsinh...

Page 9: ...the field appears darkest Thus the mechanism is adjusted Adjust the contrast Grey contrast 4 Platte swing polarizer turn until the desired contrast enters With the rotation of the polarizer over 45 ch...

Page 10: ...mission until the field appears darkest and no relief like contrasting to be seen is Prism in the intermediate ring DIK D with pin lever slightly adjust until the field appears darkest Thus the mechan...

Page 11: ...onnection with the integration ring lens supply the condensor lens darks field of maximum luminous intensity for objectives with enlargements from 25x to REAR ONE 100x ten Rear objectives with apertur...

Page 12: ...field stop polarizer with wedging screws in this situation determine Preparation again up put Tricks of the table inserting and rotation of the compensators and 4 after the requirements of the invest...

Page 13: ...indieser situation by tightening the clamping screw fix For the use of the different mf AKS mechanisms in en connection with JENAVAL you orient yourselves please in serving guidance to micro photograp...

Page 14: ...t that the eyepiece opposite Endedes of placable eyepiece unscrew eyepiece plate with line illustrated page downward into the screwed ring insert screwed ring in again into the eyepiece screw Place ca...

Page 15: ...one uses rear objectives with apertures favourably 1 with the condensor lens 0 9 although luminous intensity and resolving power of these objectives are not exhausted If its full achievement is to be...

Page 16: ...clamp fig 2 23 can the table turn be locked Note If the Y impulse of the object leader over the work area for slide 26 x is adjusted 76 mm outside Table carefully turn so that it does not fasten at th...

Page 17: ...rret It contains three fiber donates The interesting preparation place is marked by three points The points are because of the corners of a equivalent lateral triangle with 4 mm of edge length which d...

Page 18: ...f they are kept locked over night giving if into the opened pin chamber in bring and draw let some drops water Chamber thereafter one while en closed hold Otherwise cartridge replacement 28 Maintenanc...

Page 19: ...uel element fig 4 39 fix Box spanners B use fig 6 58 Table carriers fig 6 29 according to illustration 3 from the top into the swallow guidance to attack push in and with clamping screw fix box spanne...

Page 20: ...lamp tension 26 adjusting wheel for annunciator screen Fig 3 27 achromatic aplanatischer condensor lens 0 9 28 clamping screws for tables of 29 table carriers 30 clamping screw for condensor lens A T...

Page 21: ...becomes transport secured before the dispatch Before putting the Kontrasttubus KT on 45 on the stand 10 must the red marked cheese head screw trans haven protection which is at the lower surface of th...

Page 22: ......

Page 23: ......

Reviews: