DuraMax Installation site requirements
15
DuraMax with standard base
910
2478
≥
200
732
905
8
Front and side view/ M 1:30/ dimensions in millimeters
600
(500)*
600 (500)*
≥
310
1170
1360
1089
1
3
4
5
6
7
9
10
2
Top view
1 Network connection for data backup and Internet
(Teleservice)
2 2x shock-proof (Schuko) wall outlets
3 5-outlet power strip (to be provided by the customer)
4 Power cable: power supply for controller
5 Network cable: CMM - data system
6 Power cable (≥5 m): power supply for data system
(e.g. computer, printer, monitor)
7 Controller
8 Additional housing on base (only with rotary table
option)
9 Control console tray (removable)
10 Table (optional) for data system
* Minimum distance for personal safety and service
(600 mm for permanent and 500 mm for movable
walls or protective fences)
Note:
All CMM components must be accessible for
service work. All crushing hazards must be elimi-
nated.
Weight
Value
CMM
[kg]
350
CMM with standard base
[kg]
445
CMM with standard base and rotary table
[kg]
465
Workpiece, max.
[kg]
100
Floor load
Value
Mean surface pressure on the 3 support surfaces.
[N/mm
2
]
approx. 0.2
Support surfaces:
[mm
2
]
3x 10000