DuraMax Installation site requirements
12
Required connections
–
Two shock-proof (Schuko) wall outlets (max. 16 A) in the direct vicinity of the installation site. This measure must
be carried out prior to installation.
–
A 5-outlet (or more) power strip (to be provided by the customer).
Power for the CMM must be supplied by one of the wall outlets.
The power cable with corresponding plug is connected to the controller of the CMM by ZEISS service technicians dur-
ing the initial installation.
The second outlet is required for the power strip via which the data system and peripherals are powered. As general
overvoltage protection, we recommend installing a surge protector.
Connection diagram for CMM with standard base
We recommend installing the connection in accordance with the following diagram.
Note:
Comply with applicable national regulations.
L
N
PE
➅
➁
➂
➂
➆
C16A
➄
➃
➀
➊
Customer installation of local voltage (suggested)
Circuit breaker
Recommended: residual current circuit breaker
Shock-proof outlet
Connection cable (min. 5 m)
Shock-proof power strip (min. 5 outlets)
CMM
Data system and peripheral devices (max. power consumption 1000 VA)
Connection diagram for CMMs with a shopfloor base
We recommend installing the connection in accordance with the following diagram.
Note:
Comply with applicable national regulations.
L
N
PE
C16A
Customer installation of local voltage (suggested)
Circuit breaker
Recommended: residual current circuit breaker
Shock-proof outlet
5 m connection cable (included with shopfloor base)
Shopfloor base
CMM
Surge protector
Shock-proof power strip (min. 6 outlets)
Data system and peripheral devices (max. power consumption 1000 VA)