DuraMax Installation site requirements

7

Shipping crates for DuraMax

The CMM assemblies are delivered securely on shipping pallets.

Shipping crates for:

Dimensions, max. [mm]

Max. gross weight (kg)

Length

Width

Height

CMM, crates, pallet

DuraMax secured

1320

1160

2100

455

Standard base

1220

820

940

130

ShopFloor base

1220

820

1120

202

Accessories

-

-

-

-

All values are maximum values. For production reasons, the dimensions of the shipping crates (length, width and
height) may deviate by as much as +50 mm.

Intermediate storage/unpacking

Store the packed CMM assemblies in a sheltered location as close as possible to the installation site.

Storage temperature: +5° to +40°C.

Risk of corrosion if protective cover missing.

—

If the machine is stored for any length of time, the protective cover must 

not

 be removed. The

cover is a component of the CMM packaging and should always be kept with it.

—

The CMM must remain inside the protective cover at its final installation site for at least 24
hours before the cover is opened or removed. This is the only way to prevent rust damage due
to condensation.

—

Only a ZEISS service technician or specially trained personnel are authorized to remove the pro-
tective cover.

NOTE

Packing materials such as disposable transport packaging, covers or Styrofoam chips are reused by
ZEISS and can be returned to ZEISS.

ZEISS recommends storing the disposable packaging and the transport locks in a safe location (to
protect them against damage) in case the CMM has to be moved to another site at a later date.

Adaptation to room temperature

No later than 2 days before the arrival of our system specialists, bring the CMM assemblies to a storage location
with an environmental temperature of at least +15°C.

Relocating an installed CMM

 WARNING

Risk of injury if the CMM falls over due to a higher center of gravity.

Crushing of body parts

Irreparable damage to the CMM.

—

Due to the high center of gravity, the base must not be transported with the CMM mounted.
The base and CMM must always be transported separately.

Summary of Contents for DuraMax

Page 1: ...DuraMax Installation site requirements Coordinate measuring machine CMM...

Page 2: ...DuraMax Installation site requirements 2020 12 01 61221 3011202...

Page 3: ...n 9 General protection measures 9 Doorway dimensions in the measuring lab 9 Room height of the measuring lab 9 Floor 9 Floor plan 9 Floor load 9 Vibrations at the installation site 9 Environmental con...

Page 4: ...standard base 16 DuraMax with ShopFloor base 17 Securing the shopfloor base 18 DuraMax measuring table with threaded holes 19 Limit curves of permissible floor vibrations at the installation site 20...

Page 5: ...s comprised of the following components Coordinate measuring machine CMM Controller integrated Control console Base optional Data system and peripheral devices optional Note on data systems NOTE Data...

Page 6: ...lifted below the center of gravity label the cargo may tip over and be ir reparably damaged The center of gravity of the packaged CMM is located outside of the center of the shipping pallet and is lab...

Page 7: ...ctive cover at its final installation site for at least 24 hours before the cover is opened or removed This is the only way to prevent rust damage due to condensation Only a ZEISS service technician o...

Page 8: ...located when properly secured with transport braces and under the supervision of our system specialists Preparatory measures The measuring table must be clear Remove probe The control console and the...

Page 9: ...pFloor base for the dimensions and information Floor plan If you create a floor plan on a CAD system you can produce a DXF and or a DWG file containing the dimensions If you require further assistance...

Page 10: ...d in VDI VDE 2627 Environmental conditions for operational readiness The following conditions must be met to properly operate the coordinate measuring machine Permissible humidity without condensation...

Page 11: ...EMC The CMM is a class A device and may cause radio interference in living areas In such cases opera tors may be required to take adequate measures at their own expense For peripheral devices e g comp...

Page 12: ...recommend installing the connection in accordance with the following diagram Note Comply with applicable national regulations L N PE C16A Customer installation of local voltage suggested Circuit brea...

Page 13: ...equires an Internet connection An Internet connection is not required for the installation of the CMM Network topology System components are networked during start up by a ZEISS service engineer or a...

Page 14: ...e g distance to the wall and required connections can be seen in the drawing on the next page Weight Value CMM kg 350 Workpiece max kg 100 Floor load Mean pressure on the support surfaces Please chec...

Page 15: ...power supply for data system e g computer printer monitor 7 Controller 8 Additional housing on base only with rotary table option 9 Control console tray removable 10 Table optional for data system Mi...

Page 16: ...plug screws are included with delivery to anchor the standard base to the floor foundation The plug screws must be anchored into the foundation There is a risk of tipping if not properly anchored If...

Page 17: ...or monitor keyboard and mouse 6 Control console tray removable 7 Controller 8 Additional housing on base only with rotary table option Minimum distance for personal safety and service 600 mm for perma...

Page 18: ...965 Position of the boreholes for the shopfloor base top view M 1 10 dimensions in millimeters Three fastening brackets 1 and three plug screws are included with delivery for fastening the shopfloor...

Page 19: ...allation site requirements 19 DuraMax measuring table with threaded holes 494 x 494 0 100 0 1 100 0 1 100 0 1 100 0 1 200 0 1 200 0 1 200 0 1 200 0 1 0 M10 25x Measuring table M 1 4 all dimensions in...

Page 20: ...s the limit values must be compared with the square mean RMS value of the measured floor vibrations If there are any pulse shaped floor vibrations e g caused by a nearby press the limit values must be...

Page 21: ...DuraMax Installation site requirements 21...

Page 22: ...DuraMax Installation site requirements 2020 12 01 61221 3011202...

Reviews: