Installation of the AxioCam MR
AxioCam MR Instal Reference
7
jçìåíáåÖ=íÜÉ=Å~ãÉê~=çå=íÜÉ=ãáÅêçëÅçéÉ=
Please use a 1.0x or a 0.63x C-Mount adapter to mount the camera on the
microscope's TV port.
Possible adapters (example):
Port 44
Port 60
1.0 x
No. 452995 (44 C2/3" 1.0x)
No. 456105 (60 C2/3" 1.0x)
0.63 x
No. 452997 (44 C2/3" 0.63x)
No. 1069-414 (60 C2/3" 0.63x)
For the Axiovert 25 please use adapter 451268.
The adapters are not supplied with the camera.
For mounting on the microscope please remove the protection cap from the
camera's C-Mount interface. Screw in the adapter until you reach the end of
the thread. Then put the camera on the microscope's TV port. Please ensure
that no dust drops into the camera's opening.
Never touch the surface of the IR barrier filter
or the CCD sensor
with your fingers. If you remove the camera from the microscope
please close the C-Mount opening immediately using the
protection cap supplied in order to avoid dust contamination.
`çååÉÅíáåÖ=Å~ãÉê~=íç=m`f=áåíÉêÑ~ÅÉ=Äç~êÇ=
Plug in the data cable until you hear an audible click.
Place the data cable in a suitable way from the camera to the PCI
interface. Protect all cables from the microscope's hot light sources.
Plug in the data cable of the camera on PC side in the interface's
connector and not in an network connector. The interface board is
labeled with "AxioCam MR Interface".