ZEISS
6 On Screen Display (OSD) Menu | 6.3 Settings menu
1
2
Fig. 14: Bright Pixel Correction menu
No.
Name
Description
1
Correct
button
Initiates Bright Pixel Correction.
2
Reset
button
Restores the factory default Bright Pixel Correction con-
figuration.
Applying Bright Pixel Correction
Prerequisite
ü
The C-mount port is closed so that no light will reach the image sensor during the procedure.
You can close the C-mount port by either closing the light path of the microscope stand or
detaching the camera from the stand and protect sensor from light by screwing the protective
cap onto the camera's c-mount thread.
Procedure
1. Tap the
Correct
button.
à
Do not operate the camera during the procedure.
à
If light reaches the sensor, the procedure fails.
Either cover the camera properly or reset the Bright Pixel configuration before restarting
the procedure.
9
The Bright Pixel Correction procedure is applied.
The camera reboots after a successful procedure.
Instruction Manual ZEISS Axiocam 208 color / 202 mono | en-US | Rev. 4 | 426570-7011-001
37