Parts Removal/Replacement
ATLAS Model 9000
Confidential and Proprietary
Service Manual
3-10
63041A0208
3.8 Power Supply (P/S) Removal
1)
Remove the Operator panel (Section 3.3).
2)
Remove the power input cable (Patient side of P/S).
3)
Disconnect the distribution harness (P/S to Motherboard).
4)
Remove the 2 screws from the distribution harness side.
5)
Slide the module toward the front of the instrument to release.
Note –
The Motherboard Power Connector Support Tool is required when reconnecting
the power connector to the Motherboard. Place the support tool underneath the
Motherboard at the middle of the power connector to keep the Motherboard from
flexing while attaching the power connector.
Figure 3.9 Power Supply
Power
Connector
Support Tool
Motherboard
Power Connector