4.1 Introduction
This section contains the procedures for adjustment and calibration of the ATLAS Model
9000. The calibration on the ATLAS 9000 is limited to verification. Hardware calibration is
not possible in the field due to the complexity and sensitivity of the alignments.
The procedures in this section assume that the Field Support Engineer is familiar with the
operation of the instrument. Complete operating instructions are contained in the ATLAS
User Manual.
4.2 Calibration Verification Check
Purpose
To determine whether the ATLAS Model 9000 is in Calibration.
Special Tools / Equipment Required
•
ATLAS Test Eye Tool
Procedure
1) Switch on the instrument and let it power up to the main menu.
2) Insert the ATLAS Test Eye Tool onto the Chinrest Assembly.
3) Click on
Tools > Calibration Check
from the toolbar.
4) Follow the instructions on-screen to complete the calibration. Once all test spheres have
been read, a Results Calibration Check Screen (Figure 4.1) will be displayed indicating
your status.
Summary of Contents for ATLAS 9000
Page 4: ......
Page 49: ...Diagrams ATLAS Model 9000 Confidential and Proprietary Service Manual 6 2 63041A0208 Notes...
Page 51: ...FIGURE 6 2 ATLAS Model 9000 System Wiring Diagram...
Page 53: ...Parts ATLAS Model 9000 Confidential and Proprietary Service Manual 7 2 63041A0208 Notes...
Page 76: ...Appendix A ATLAS Model 9000 Confidential and Proprietary Service Manual A 4 63041A0208...
Page 112: ...Appendix D ATLAS Model 9000 Confidential and Proprietary Service Manual D 2 63041A0208...
Page 114: ...Appendix E ATLAS Model 9000 Confidential and Proprietary Service Manual E 2 63041A0208...
Page 116: ...Appendix F ATLAS Model 9000 Confidential and Proprietary Service Manual F 2 63041A0208...