Appendix B
ATLAS Model 9000
Confidential and Proprietary
Service Manual
B-14
63041A0208
B.3 System Backup: Create a New Hard Drive Image
Carl Zeiss Meditec strongly recommends using Acronis to create a new hard drive image
(called a new “archive” in Acronis) after completing each successful upgrade of the
system software, so later recovery may be to the most recent software version.
Note –
Each Acronis archive uses approximately 3.5 to 4.0 GB of hard disk space.
Therefore, we recommend that you do not make archives except after system
software upgrades. If desired, you may delete archives older than the most recent,
to make more room for exam data.
System Backup
1)
Launch Acronis as explained in Section B.2, System Recovery.
Figure B-18 Acronis Upon Launch (through Windows)