Safety Notes 

 

HBO 50 

Care, Maintenance, Disposal, Warranty 

Carl Zeiss

 

 

 
M60-2-0014  e    05/2012 

1.6

 

Notes on  

care and maintenance 

Proper function of the microscope illuminator 
requires that only components and spares are 

used that have been supplied or approved for 
use by Carl Zeiss. In case of doubt, contact Carl 

Zeiss Service. 
Do not perform any actions on the device other 

than those described in this operating manual. 
We want to emphasize that actions not 
described here may only be performed by 

authorized persons. 
Damaged equipment’s or components may only 

be repaired and serviced by Carl Zeiss Service. 
 

 

 

 

 

 

Replace the burner (lamp) after the 

mean 

lifetime

  

(

refer to data sheet of lamp manufacturer

has elapsed;  
otherwise, there is an increased 

risk of 

explosion

You can read the operating time of the lamp 

from the hour meter of the corresponding 
power supply unit (refer to Fig. 4/

2

). 

 

 

To replace the burner, follow the steps 
described in 

Section 3.3

. 

 

When replacing the burner, check the leads 

of the HBO 50. 

If the housing of the microscope illuminator is 

dirty, you can clean it. 
Do not use a solvent for cleaning. 

Immerse the cleaning cloth in a mixture of 
water and a dash of a detergent, remove the 

dirt and dab the housing dry with a clean cloth. 

To clean the lamp power supply unit, only use a 
dry, lintless cloth taking care that no dust or 

other foreign bodies enter the device through 

the ventilation slots. 
In the case of damages (also external ones), 
immediately shut down the device and have it 

serviced. 
 
 

1.7

 

Notes on disposal 

 

The HBO 50 AC (L1 / L2) burners contain little 

quantities of mercury! 
Do not dispose of the burners with household 
waste. 

Dispose of the lamps in their original packing in 

compliance with legislation. 
In case of doubt, contact Carl Zeiss Service or an 
OSRAM agency. 
 
 

1.8

 

Notes on warranty 

The manufacturer of the equipment is not liable 

for damages caused by unauthorized 
interventions, particularly those caused by the 

removal or the exchange of device components 
or by the use of accessory items of other 
manufacturers. Unauthorized intervention 

will 

void all warranties

.

 

 

Summary of Contents for 423003-9000

Page 1: ...Operating manual HBO 50 Microscope Illuminator...

Page 2: ...nd communication of its contents are not permitted Violations will entail an obligation to pay compensation All rights reserved in the event of granting of patents or registration of an utility model...

Page 3: ...Intended use 8 2 2 Components 8 2 3 Technical data 9 3 Start up 11 3 1 Inserting and replacing the burner 11 3 2 Starting up the ebq 50 ac lamp power supply 13 3 3 Focusing and centering the burner 1...

Page 4: ...ebq 52 ac lamp power supply Particularly observe the safety notes given in this operating manual For the handling and operation of the burner lamp of the HBO 50 observe the specifications and safety i...

Page 5: ...liance with the WEEE Directive 2002 96 EC The equipment carries the mark The equipment complies with the standards DIN EN 61326 Class B and DIN EN 6100 3 2 1 4 Notes on unpacking transport and storage...

Page 6: ...erating room is supplied with fresh air or ozone is conducted into the open air through an exhaust fan Only authorized and instructed personnel or Carl Zeiss Service are allowed to replace and align t...

Page 7: ...Section 3 3 When replacing the burner check the leads of the HBO 50 If the housing of the microscope illuminator is dirty you can clean it Do not use a solvent for cleaning Immerse the cleaning cloth...

Page 8: ...0 Microscope Illuminator consists of the following components Fig 1 Lamp housing 1 1 with mount 1 6 for connection to the microscope and removable lamp compartment Permanently mounted collector Perman...

Page 9: ...90 VA Fuses 2x T1 6A breaking capacity H manufacturer LITTLEFUSE series 215 Protection Class I Protection Type IP 20 Output power 50 W 3 W Lamp voltage 34 V to 55 V suitable for L1 and L2 Burner Mercu...

Page 10: ...Illuminator Carl Zeiss Technical Data HBO 50 10 M60 2 0014 e 05 2012 Fig 2 Line spectrum of HBO 50 mercury vapor short arc lamp 0 10 20 30 40 50 60 70 80 90 100 200 300 400 500 600 700 800 Wavelength...

Page 11: ...result in contacting faults due to strong heat development in use Not remove or damage the cable for the cooling element Avoid fingerprints on glass components of the burner if necessary remove them...

Page 12: ...spring lever Press spring lever 3 5 for lamp socket 3 4 down once more and insert burner into lamp socket Do not touch glass bulb of burner Let go of spring lever Press spring lever 3 6 for cooling el...

Page 13: ...Power switch 2 Hour meter 3 Power input connector with fuse F2 4 Connector for lamp plug Fig 4 ebq 50 ac lamp power supply When the device is connected to the mains the full operating or starting volt...

Page 14: ...ower switch 4 2 on the front panel of the lamp power supply the burner starts automatically The HBO 50 microscope illuminator is to be adjusted in two steps coarse adjustment without adjusting aid and...

Page 15: ...scription given in the operating manual of the microscope Pull out the adjusting aid of the upper part of the microscope 8 1 The brighter cathode spot of the lamp and its slightly darker mirror image...

Page 16: ...ower connector checked by authorized persons only Power indicator in power switch lights in switch position I No function of hour meter Check fuse F 1 Disconnect power cable from device Make fuse acce...

Reviews: