Start-up
Microscope Illuminator
Carl Zeiss
Focusing and centering the lamp
HBO 50
14
M60-2-0014 e 05/2012
3.3
Focusing and centering the
burner
1 Screw for axial lamp adjustment
2 Screw for vertical lamp adjustment
3 Screw for lateral lamp adjustment
Fig. 5
Adjusting controls of microscope
illuminator
•
Switch on the microscope illuminator with
the power switch (4/
2
) on the front panel of
the lamp power supply; the burner starts
automatically.
The HBO 50 microscope illuminator is to be
adjusted in two steps: coarse adjustment
(without adjusting aid) and subsequent fine
adjustment by means of the adjusting aid on
the microscope.
Adjustment without adjusting aid
Wear UV protective eyewear in adjustment
without adjusting aid to protect the eyes from
exposure to UV radiation!
•
Direct the light beam onto a projection
surface (e.g. bright wall) that is at least 3 m
away.
Fig. 6
Cathode spots before adjustment