Start-up 

Microscope Illuminator 

Carl Zeiss 

Focusing and centering the lamp 

HBO 50 

 

 

14 

M60-2-0014  e    05/2012 

3.3

 

Focusing and centering the 

burner 

 

 

1  Screw for axial lamp adjustment 

2  Screw for vertical lamp adjustment  
3  Screw for lateral lamp adjustment 

Fig. 5 

Adjusting controls of microscope 
illuminator 

 

 

Switch on the microscope illuminator with 
the power switch (4/

2

) on the front panel of 

the lamp power supply; the burner starts 

automatically.

 

 

The HBO 50 microscope illuminator is to be 

adjusted in two steps: coarse adjustment 
(without adjusting aid) and subsequent fine 

adjustment by means of the adjusting aid on 
the microscope. 
 

Adjustment without adjusting aid 

 

Wear UV protective eyewear in adjustment 

without adjusting aid to protect the eyes from 
exposure to UV radiation! 

 

Direct the light beam onto a projection 

surface (e.g. bright wall) that is at least 3 m 
away. 

 

 

Fig. 6 

Cathode spots before adjustment 

Summary of Contents for 423003-9000

Page 1: ...Operating manual HBO 50 Microscope Illuminator...

Page 2: ...nd communication of its contents are not permitted Violations will entail an obligation to pay compensation All rights reserved in the event of granting of patents or registration of an utility model...

Page 3: ...Intended use 8 2 2 Components 8 2 3 Technical data 9 3 Start up 11 3 1 Inserting and replacing the burner 11 3 2 Starting up the ebq 50 ac lamp power supply 13 3 3 Focusing and centering the burner 1...

Page 4: ...ebq 52 ac lamp power supply Particularly observe the safety notes given in this operating manual For the handling and operation of the burner lamp of the HBO 50 observe the specifications and safety i...

Page 5: ...liance with the WEEE Directive 2002 96 EC The equipment carries the mark The equipment complies with the standards DIN EN 61326 Class B and DIN EN 6100 3 2 1 4 Notes on unpacking transport and storage...

Page 6: ...erating room is supplied with fresh air or ozone is conducted into the open air through an exhaust fan Only authorized and instructed personnel or Carl Zeiss Service are allowed to replace and align t...

Page 7: ...Section 3 3 When replacing the burner check the leads of the HBO 50 If the housing of the microscope illuminator is dirty you can clean it Do not use a solvent for cleaning Immerse the cleaning cloth...

Page 8: ...0 Microscope Illuminator consists of the following components Fig 1 Lamp housing 1 1 with mount 1 6 for connection to the microscope and removable lamp compartment Permanently mounted collector Perman...

Page 9: ...90 VA Fuses 2x T1 6A breaking capacity H manufacturer LITTLEFUSE series 215 Protection Class I Protection Type IP 20 Output power 50 W 3 W Lamp voltage 34 V to 55 V suitable for L1 and L2 Burner Mercu...

Page 10: ...Illuminator Carl Zeiss Technical Data HBO 50 10 M60 2 0014 e 05 2012 Fig 2 Line spectrum of HBO 50 mercury vapor short arc lamp 0 10 20 30 40 50 60 70 80 90 100 200 300 400 500 600 700 800 Wavelength...

Page 11: ...result in contacting faults due to strong heat development in use Not remove or damage the cable for the cooling element Avoid fingerprints on glass components of the burner if necessary remove them...

Page 12: ...spring lever Press spring lever 3 5 for lamp socket 3 4 down once more and insert burner into lamp socket Do not touch glass bulb of burner Let go of spring lever Press spring lever 3 6 for cooling el...

Page 13: ...Power switch 2 Hour meter 3 Power input connector with fuse F2 4 Connector for lamp plug Fig 4 ebq 50 ac lamp power supply When the device is connected to the mains the full operating or starting volt...

Page 14: ...ower switch 4 2 on the front panel of the lamp power supply the burner starts automatically The HBO 50 microscope illuminator is to be adjusted in two steps coarse adjustment without adjusting aid and...

Page 15: ...scription given in the operating manual of the microscope Pull out the adjusting aid of the upper part of the microscope 8 1 The brighter cathode spot of the lamp and its slightly darker mirror image...

Page 16: ...ower connector checked by authorized persons only Power indicator in power switch lights in switch position I No function of hour meter Check fuse F 1 Disconnect power cable from device Make fuse acce...

Reviews: