Microscope Illuminator 

Description 

HBO 50 

Technical Data 

Carl Zeiss 

 

 

M60-2-0014  e    05/2012 

9  

2.3

 

Technical data 

Lamp housing 

HBO 50 Illuminator 

Cat. No. 423003-9000 and 423003-9010

 

Length x Width x Height 

235 mm x 165 mm x 86 mm 

Connector to microscope stand 

Mount 

Connection to lamp power supply  Cable with special plug 

Lamp power supply 

ebq 52 ac lamp power supply for HBO 50 

Cat. No. 432607-0000-000

 

Power requirements 

100 to 240 V AC ±10 % 

Line frequency 

50 / 60 Hz; externally switchable 

Power consumption 

Max. 90 VA 

Fuses 

2x T1.6A; breaking capacity H; manufacturer: LITTLEFUSE, series 215 

Protection Class 

Protection Type 

IP 20 

Output power  

50 W (±

 

3 W) 

Lamp voltage 

34 V to 55 V (suitable for L1 and L2) 

Burner 

Mercury-vapor short-arc lamp 

For technical data of burners,  

refer to the specifications of the 
manufacturers. 
Recommended for use: 

OSRAM

  

Type HBO 50 W/AC / L1 or HBO 50 W/AC / L2 
acc. to Cat. No. 381619-0000-000 

Lamp voltage 

L1: 39 ... 45 V / L2: 34 ... 39 V 

Lamp current 

L1: 1.30 A / L2: 1.45 A 

Power 

50 W 

Mean luminous flux 

2,000 lm 

Luminous area 

0.3 mm x 1 mm 

Average service life of lamp 

Type HBO 50 W/AC / L1; L2 

100 h  

Ambient conditions 

For indoor use only 

Permissible temperature range 

 5 ... 40 °C 

Relative humidity 

< 80 %, no condensation 

 

 

 

Summary of Contents for 423003-9000

Page 1: ...Operating manual HBO 50 Microscope Illuminator...

Page 2: ...nd communication of its contents are not permitted Violations will entail an obligation to pay compensation All rights reserved in the event of granting of patents or registration of an utility model...

Page 3: ...Intended use 8 2 2 Components 8 2 3 Technical data 9 3 Start up 11 3 1 Inserting and replacing the burner 11 3 2 Starting up the ebq 50 ac lamp power supply 13 3 3 Focusing and centering the burner 1...

Page 4: ...ebq 52 ac lamp power supply Particularly observe the safety notes given in this operating manual For the handling and operation of the burner lamp of the HBO 50 observe the specifications and safety i...

Page 5: ...liance with the WEEE Directive 2002 96 EC The equipment carries the mark The equipment complies with the standards DIN EN 61326 Class B and DIN EN 6100 3 2 1 4 Notes on unpacking transport and storage...

Page 6: ...erating room is supplied with fresh air or ozone is conducted into the open air through an exhaust fan Only authorized and instructed personnel or Carl Zeiss Service are allowed to replace and align t...

Page 7: ...Section 3 3 When replacing the burner check the leads of the HBO 50 If the housing of the microscope illuminator is dirty you can clean it Do not use a solvent for cleaning Immerse the cleaning cloth...

Page 8: ...0 Microscope Illuminator consists of the following components Fig 1 Lamp housing 1 1 with mount 1 6 for connection to the microscope and removable lamp compartment Permanently mounted collector Perman...

Page 9: ...90 VA Fuses 2x T1 6A breaking capacity H manufacturer LITTLEFUSE series 215 Protection Class I Protection Type IP 20 Output power 50 W 3 W Lamp voltage 34 V to 55 V suitable for L1 and L2 Burner Mercu...

Page 10: ...Illuminator Carl Zeiss Technical Data HBO 50 10 M60 2 0014 e 05 2012 Fig 2 Line spectrum of HBO 50 mercury vapor short arc lamp 0 10 20 30 40 50 60 70 80 90 100 200 300 400 500 600 700 800 Wavelength...

Page 11: ...result in contacting faults due to strong heat development in use Not remove or damage the cable for the cooling element Avoid fingerprints on glass components of the burner if necessary remove them...

Page 12: ...spring lever Press spring lever 3 5 for lamp socket 3 4 down once more and insert burner into lamp socket Do not touch glass bulb of burner Let go of spring lever Press spring lever 3 6 for cooling el...

Page 13: ...Power switch 2 Hour meter 3 Power input connector with fuse F2 4 Connector for lamp plug Fig 4 ebq 50 ac lamp power supply When the device is connected to the mains the full operating or starting volt...

Page 14: ...ower switch 4 2 on the front panel of the lamp power supply the burner starts automatically The HBO 50 microscope illuminator is to be adjusted in two steps coarse adjustment without adjusting aid and...

Page 15: ...scription given in the operating manual of the microscope Pull out the adjusting aid of the upper part of the microscope 8 1 The brighter cathode spot of the lamp and its slightly darker mirror image...

Page 16: ...ower connector checked by authorized persons only Power indicator in power switch lights in switch position I No function of hour meter Check fuse F 1 Disconnect power cable from device Make fuse acce...

Reviews: