AC Mains Connection
The ZyPer4K-M Encoder and Decoder units are powered through an external medical-grade
AC/DC converter unit.
The Encoder and Decoder units must be powered and connected using the supplied medical
grade external AC/DC converter. The voltage supplied to the Encoder and Decoder units from
the external supply is 12VDC at less than 2.0A
•
Do not substitute the supplied AC/DC power supply unit with another component.
The overall system compliance is dependent on each of the provided elements
being present and operational.
INFORMATION: The introduction of a foreign, non-ZeeVee element nullifies
any claim of regulatory compliance asserted by ZeeVee.
The AC/DC power supply shall be powered from the appropriately rated AC mains
voltage which allows for a wide range of input, generally from 100VAC through 240VAC.
Only use a power cord with an appropriately rated protective earth connection. The protective
earth connection is essential to the overall safety of the system.
WARNING: Bypassing or omitting the protective earth
connection increases the risk of electrical shock and
radio emissions
Install the equipment in such a manner as the IEC power cord ingress in to the AC/DC adapter
is accessible as this is the boundary of the AC mains system and the point at which the
ZyPer4K-M equipment is disconnected from the AC mains power if needed.
The equipment should be installed as near as is practicable to an AC outlet.
•
Do not substitute or extend the DC cable from the AC/DC unit.
The ZyPer4K system is intended and rated for continuous operation