Sample chamber overflows through the injection port.
Sample chamber overflows through the injection port.
Sample chamber overflows through the injection port.
Sample chamber overflows through the injection port.
•
POSSIBLE CAUSE: Waste tubing line in pumps crimped or stuck shut.
•
CORRECTIVE ACTION: Disassemble pumps and open tubing or replace if
more than 6 months since last replaced.
•
•
POSSIBLE CAUSE: Waste line blocked.
•
CORRECTIVE ACTION: Clear tubing or replace.
•
•
POSSIBLE CAUSE: Waste line check valve stuck open.
•
CORRECTIVE ACTION: Flush check valve or replace.
•
POSSIBLE CAUSE: Bottle vent plugged, causing positive pressure in the waste
bottle.
•
CORRECTIVE ACTION: Loosen waste bottle cap to verify, clear blockage or
replace vent.
•
POSSIBLE CAUSE: Capillary tube plug broken loose from tube and stuck in
waste chamber port.
•
CORRECTIVE ACTION: Disassemble chamber, check for blockage in outlet
port (upper).
•
•
POSSIBLE CAUSE: Sample chamber outlet blocked.
•
CORRECTIVE ACTION: Disassemble chamber, check for blockage in outlet
tube due to salt buildup, and clean if necessary.
•
Stir bar fails to spin.
Stir bar fails to spin.
Stir bar fails to spin.
Stir bar fails to spin.
•
POSSIBLE CAUSE: Stir bar stuck to chamber.
•
CORRECTIVE ACTION: Clean the sample chamber.
•
SECTION: 8.3
Summary of Contents for 1500 Sport
Page 2: ......
Page 8: ...1 1 1 1 2 2 2 2 ...
Page 46: ...5 5 5 5 1 1 1 12 2 2 2 ...
Page 68: ...6 6 6 6 22 22 22 22 ...
Page 78: ...7 7 7 7 10 10 10 10 ...
Page 86: ......
Page 100: ...9 9 9 9 14 14 14 14 ...
Page 114: ...12 12 12 12 4 4 4 4 ...
Page 116: ...13 13 13 13 2 2 2 2 ...
Page 122: ...14 14 14 14 6 6 6 6 ...
Page 124: ...15 15 15 15 2 2 2 2 ...
Page 126: ...16 16 16 16 2 2 2 2 ...
Page 128: ...17 17 17 17 2 2 2 2 ...
Page 132: ...18 18 18 18 4 4 4 4 ...